Reports enhanced dopamine detection in a novel microsystem, suggesting improved neurochemical analysis techniques.
This paper reports on a new neurochemical recording microsystem that employs 400V/s fast-scan cyclic voltammetry (FSCV) at a carbon-fiber microelectrode (CFM). Featuring analog background current subtraction in situ to reduce the requisite input dynamic range of the sensing front-end, this design employs a 1 st -order, continuous-time, delta-sigma modulator (ΔΣM) with 1b quantization clocked at a frequency of 1.6MHz. The analog building blocks of the microsystem are fabricated in TSMC 0.18μm 1P/6M CMOS, while the digital blocks such as memories, decimation filter, and control units are implemented on a field-programmable gate array (FPGA). Experimental results from flow injection analysis (FIA) tests show a highly linear response (r 2 = 0.988) in calibration curve for dopamine with a measured sensitivity of ~35.9nA/μM for dopamine concentration levels in the range of 0-1μM.
No takes yet. Share an insight, caveat, or question.
Zamani et al. (2020) studied this question.