Key result
Real time three-dimensional echocardiography derived maximal left atrial volume showed high correlation with CMR (r=0.93) and no significant difference (86 vs 89 mL, P=0.079).
Why the study?
The accuracy of real time three-dimensional echocardiography (RT-3DE) for evaluating left atrial volume against CMR had not been reported in heart transplant recipients.
Does real time three-dimensional echocardiography accurately quantify left atrial volume compared to cardiac magnetic resonance in orthotopic heart transplant recipients?
Cross-Sectional (n=31)
Does real time three-dimensional echocardiography accurately quantify left atrial volume compared to cardiac magnetic resonance in orthotopic heart transplant recipients?
Effect estimate: r = 0.93
p-value: p=0.079
RT-3DE is a more accurate alternative to 2DE and comparable to CMR for quantifying maximal left atrial volume in heart transplant recipients.
May support RT-3DE over 2DE for maximal LAV post-transplant; leaves open larger prospective validation versus CMR.
INTRODUCTION: The accuracy of real time three-dimensional echocardiography (RT-3DE) in evaluating left atrial volume (LAV) of heart transplant recipients against cardiac magnetic resonance (CMR) has not been reported. The aim of this study was to compare LAV with RT-3DE with respect to CMR in heart transplant recipients. METHODS: Thirty-one heart transplant recipients who received echocardiogram and CMR examination on the same day were prospectively enrolled. The maximal LAV, minimal LAV by RT-3DE, and two-dimensional echocardiography (2DE) were compared with CMR measurements. Inter-technique comparisons included Pearson's correlation coefficient and Bland-Altman analysis. Reproducibility of 2DE and RT-3DE technique was assessed by intra-class correlation coefficient (ICC). RESULTS: RT-3DE-derived LAV values showed higher correlation with CMR than 2DE measurements in heart transplant recipients (r = .93 vs r = .76 for maximal LAV; r = .91 vs r = .81 for minimal LAV). Two-dimensional echocardiography underestimated maximal LAV by 10 ± 31 mL and minimal LAV by 26 ± 26 mL. Although RT-3DE underestimated minimal LAV 15 ± 19 mL, no significant difference between RT-3DE and CMR was observed in maximal LAV (RT-3DE: 86 ± 22 mL; CMR: 89 ± 23 mL, P = .079), with a negligible bias of 3 mL. Inter-observer and intra-observer agreement were excellent for 2DE and RT-3DE parameters. CONCLUSION: Compared with CMR reference, RT-3DE-derived LAV measurements are more accurate than 2DE-based analysis in heart transplant recipients, especially with regard to the assessment of maximal LAV. RT-3DE may be a valid alternative to CMR for quantification LAV in heart transplant recipients.
No takes yet. Share an insight, caveat, or question.
Zhu et al. (2020) conducted a cross-sectional in Heart transplant recipients (n=31). Real time three-dimensional echocardiography (RT-3DE) vs. Cardiac magnetic resonance (CMR) and two-dimensional echocardiography (2DE) was evaluated on Maximal left atrial volume (LAV) (r = 0.93, p=0.079). Real time three-dimensional echocardiography derived maximal left atrial volume showed high correlation with CMR (r=0.93) and no significant difference (86 vs 89 mL, P=0.079).