Key result
Site-selective pacing using the Model 3830 lead in patients with congenital heart disease achieved 100% successful pacing with a 2.9% lead displacement rate over a median 21.8 months.
Why the study?
Is the Model 3830 lumenless lead system safe and effective for pacing in patients with congenital heart disease?
Cohort (n=90)
Yes
Is the Model 3830 lumenless lead system safe and effective for pacing in patients with congenital heart disease?
The Model 3830 lumenless lead system is safe and effective for pacing in patients with congenital heart disease, demonstrating acceptable complication and displacement rates in a technically challenging population.
May support Model 3830 use for pacing in congenital heart disease; leaves open need for randomized confirmation of long-term outcomes.
Background: Long‐term pacing is frequently necessary in patients with congenital heart disease (CHD). Preservation of ventricular function and avoidance of venous occlusion is important in these patients. Site‐selective pacing with a smaller diameter lead is achievable with the model 3830 lead (SelectSecure®, Medtronic Inc., Minneapolis, MN, USA), which was specifically designed to target these complications. We describe our initial experience with the Model 3830 lead in patients with CHD. Methods: Retrospective analysis of all patients undergoing site‐selective implantation of a Model 3830 lead(s) from two congenital heart centers (Bristol, UK, and Dublin, Ireland) from October 2004 until February 2008. Results: We implanted 139 SelectSecure® leads (atrial n = 70; ventricular n = 69) in 90 patients (57 male) with CHD. Median age at implantation: 13.4 years (1.1–59.2 years), median weight: 43 kg. Sixty‐nine patients (76%) were children (<18 years). Indications for lead implantation included atrioventricular block (n = 55), sinus node disease (n = 18), implantable cardiac defibrillator (n = 12), antitachycardia pacing (n = 4), and cardiac resynchronization (n = 1). Twenty‐two patients underwent pre‐existing lead extraction during the same procedure. All the attempted procedures resulted in successful pacing. One patient had a significantly raised threshold at implantation. There was no procedural mortality. There were two procedural complications. Three patients required lead repositioning for increasing thresholds early postprocedure (<6 weeks). Four leads (2.9%) had displaced on median follow‐up of 21.8 months (0.5–42 months). Conclusions: The Model 3830 lead is safe and effective in patients with CHD. This is a technically challenging patient group yet procedural complication and lead displacement rates are acceptable.
No takes yet. Share an insight, caveat, or question.
Chakrabarti et al. (2009) conducted a cohort in Congenital heart disease (n=90). Model 3830 lead (SelectSecure) was evaluated on Successful pacing. Site-selective pacing using the Model 3830 lead in patients with congenital heart disease achieved 100% successful pacing with a 2.9% lead displacement rate over a median 21.8 months.