Key result
Cavotricuspid isthmus ablation with a gold-tip catheter eliminated the need for an irrigated-tip catheter compared to a platinum-iridium tip (0% vs 25%, P<0.001) and reduced procedure time.
Why the study?
Does cavotricuspid isthmus ablation with an 8-mm gold-tip catheter improve procedural success and reduce procedure time compared to a platinum-iridium tip catheter in patients undergoing CTI ablation?
Does cavotricuspid isthmus ablation with an 8-mm gold-tip catheter improve procedural success and reduce procedure time compared to a platinum-iridium tip catheter in patients undergoing CTI ablation?
Gold-tip catheters for CTI ablation using the maximum voltage-guided technique significantly reduce procedure time, fluoroscopy time, and the number of RF applications compared to platinum-iridium catheters.
Answers are grounded in this paper and cite their sources.
Gold-tip catheters may reduce irrigated-tip needs and shorten CTI ablation; leaves open whether randomized trials confirm superiority.
BACKGROUND: Gold has excellent electrical conductive properties and creates deeper and wider lesions than platinum-iridium during radiofrequency (RF) ablation in vitro. We tested the maximum voltage-guided technique (MVGT) of cavotricuspid isthmus (CTI) ablation using two 8-mm tip catheters containing gold (group G) or platinum-iridium (group PI). METHODS: We enrolled 31 patients who underwent CTI ablation. In group G (n = 15) CTI ablation was performed with a gold-tip ablation catheter, while in group PI (n = 16) a platinum-iridium tip was used. Ablation was guided by CTI potentials with the highest amplitude until achievement of bidirectional block (BIB). If BIB was not achieved after 10 RF applications, RF was delivered via a 3.5-mm irrigated-tip catheter. Success rate, procedure duration, duration of fluoroscopic exposure, and number of RF applications were measured. RESULTS: BIB was achieved in all patients in group G, while in group PI an irrigated tip was used in four patients (0% vs 25%, P < 0.001). These four patients required a total of 21 additional RF applications (5.25 +/- 2.22). Procedure time (56.4 +/- 12 vs 73.1 +/- 15 minutes P < 0.05) and fluoroscopic explosure (4.9 +/- 2.3 vs 7.1 +/- 3.8 minutes, P < 0.01) were shorter in group G than in group PI. Mean number of RF applications was lower (4.6 +/- 1.9 vs 6.6 +/- 3.1 P < 0.001) and total RF duration shorter (280 +/- 117 vs 480 +/- 310 seconds) in group G than in group PI. No difference was observed in the number of recurrences at a 6 month-follow up (1 in group G vs 1 in group PI). CONCLUSION: Using the MVGT of CTI ablation, gold-tip catheters were associated with shorter procedural and fluoroscopic times, and fewer RF applications.
No takes yet. Share an insight, caveat, or question.
Kardos et al. (2009) studied Cavotricuspid isthmus (CTI) ablation (n=31). Gold-tip ablation catheter (8-mm) vs. Platinum-iridium tip ablation catheter (8-mm) was evaluated on Need for an irrigated-tip catheter to achieve bidirectional block (p=<0.001). Cavotricuspid isthmus ablation with a gold-tip catheter eliminated the need for an irrigated-tip catheter compared to a platinum-iridium tip (0% vs 25%, P<0.001) and reduced procedure time.