Key result
Pressure wires with optical fibers yielded FFR values highly correlated with conventional wires (r = 0.81, P < 0.001), with no significant difference in mean values (0.83 vs 0.83, P = 0.66).
Why the study?
Although newly developed pressure wires with optical fibers are available for FFR measurement, their safety and accuracy are not clear.
Does FFR measurement using pressure wires with optical fibers provide comparable accuracy to conventional FFR wires in patients with intermediate coronary artery stenosis?
Observational (n=51)
Does FFR measurement using pressure wires with optical fibers provide comparable accuracy to conventional FFR wires in patients with intermediate coronary artery stenosis?
Effect estimate: r = 0.81
Absolute Event Rate: 0.83% vs 0.83%
p-value: p=<0.001
Optical-based FFR pressure wires are safe, reliable, and correlate excellently with conventional FFR wires for assessing intermediate coronary stenosis.
BACKGROUND: Fractional flow reserve (FFR) measurement is the gold standard for identifying the functional severity of coronary artery disease. Although we can use newly developed pressure wires with optical fibers are now available, their safety and accuracy for FFR measurement are not clear. Therefore, we planned a clinical comparison study between pressure wires with optical fibers and the conventional FFR device. METHODS: We prospectively enrolled 51 patients (51 lesions) with intermediate coronary artery stenosis. For these lesions, FFR measurements with pressure wires with optical fibers were compared with those obtained with a conventional wire. RESULTS: All procedures were successfully completed without any complications. The procedure time with pressure wires with optical fibers and a conventional wire was 6.8 ± 3.0 and 6.9 ± 2.6 minutes (P = 0.89), respectively. There was no significant difference in FFR values between pressure wires with optical fibers and a conventional wire (0.83 ± 0.10 vs. 0.83 ± 0.12, P = 0.66). An excellent correlation was observed between FFR obtained with pressure wires with optical fibers and FFR obtained with a conventional wire (r = 0.81, P < 0.001). The pressure drift before and after FFR measurements was not significantly different between pressure wires with optical fibers and a convention wire (-0.01 ± 0.02 vs. -0.02 ± 0.06, P = 0.42). There was no significant difference in pressure drift defined as >0.03 or <-0.03 between pressure wires with optical fibers and a convention wire (13.7 vs. 17.6%, P = 0.79). CONCLUSION: Excellent correlations were observed between FFR values measured using pressure wires with optical fibers and a conventional wire. Furthermore, measurement of FFR with pressure wires with optical fibers was as well tolerated and reliable as that with a conventional wire.
No takes yet. Share an insight, caveat, or question.
Saka et al. (2020) conducted an observational in Intermediate coronary artery stenosis (n=51). Pressure wires with optical fibers vs. Conventional fractional flow reserve (FFR) wire was evaluated on Fractional flow reserve (FFR) values (r = 0.81, p=<0.001). Pressure wires with optical fibers yielded FFR values highly correlated with conventional wires (r = 0.81, P < 0.001), with no significant difference in mean values (0.83 vs 0.83, P = 0.66).