Why the study?
What are the sample size requirements for clinical trials of new transcatheter replacement valves and valve repair devices?
What are the sample size requirements for clinical trials of new transcatheter replacement valves and valve repair devices?
Answers are grounded in this paper and cite their sources.
Recommends a minimum sample size of 150 patients and 400 patient-years of follow-up for clinical trials of transcatheter valve devices to facilitate regulatory assessment.
ObjectivesTo provide the scientific basis for the sample size required for clinical trials of new transcatheter replacement valves and valve repair devices.BackgroundFor regulatory approval of surgically implanted valves, linearized rates of individual types of adverse events with a new valve are compared to the average rates of the same events with previously approved valves, known as objective performance criteria (OPC). OPC are used to calculate the sample sizes necessary for regulatory submission.Methods and recommendationsOPC do not exist for transcatheter valves and valve repair devices because average adverse event rates have not yet been determined. Therefore an alternative approach was devised without the benefit of historical OPC, by calculating the number of patient-years necessary to detect adverse event rates within an “acceptable” range. For example, 400 patient-years of follow-up would detect an adverse event rate of 2.4%/patient year. For consistency with the ISO standard for surgical valves, a minimum sample size of 150 patients followed for at least 1 year and a minimum follow-up of 400 patient-years is recommended unless the manufacturer can adequately justify (using the table provided) a smaller number of follow-up years based on the expectation of a higher adverse event rate in a high risk population.ConclusionsIt is hoped that the calculated sample size requirements already incorporated into ISO 5910:2018 (valve repair devices) will also be incorporated into forthcoming revision of ISO 5840–3 (transcatheter valve replacement devices), and that they should facilitate regulatory assessment.
No takes yet. Share an insight, caveat, or question.
Butchart et al. (2018) studied Transcatheter valve replacement and valve repair. Transcatheter valve replacement and valve repair devices was evaluated. A minimum sample size of 150 patients followed for at least 1 year (400 patient-years) is recommended for clinical trials of transcatheter valve devices to detect adverse event rates.
Synapse has enriched 10 related papers. Consider them for comparative context: