Key result
The Boston Sadra Lotus valve, presence of artifacts, and higher body mass index were independent predictors of poor image quality in rotational angiography after TAVI.
Why the study?
What are the determinants of image quality of rotational angiography for on-line assessment of frame geometry after transcatheter aortic valve implantation?
Observational (n=179)
No
What are the determinants of image quality of rotational angiography for on-line assessment of frame geometry after transcatheter aortic valve implantation?
Rotational angiography with motion compensation provides good image quality for evaluating frame geometry after TAVI in most patients, though higher BMI, specific valve types (Lotus), and artifacts independently predict poor image quality.
Answers are grounded in this paper and cite their sources.
These predictors may limit rotational angiography reliability post-TAVI in select patients; leaves open whether protocol adjustments enhance frame assessment in prospective studies.
To study the determinants of image quality of rotational angiography using dedicated research prototype software for motion compensation without rapid ventricular pacing after the implantation of four commercially available catheter-based valves. Prospective observational study including 179 consecutive patients who underwent transcatheter aortic valve implantation (TAVI) with either the Medtronic CoreValve (MCS), Edward-SAPIEN Valve (ESV), Boston Sadra Lotus (BSL) or Saint-Jude Portico Valve (SJP) in whom rotational angiography (R-angio) with motion compensation 3D image reconstruction was performed. Image quality was evaluated from grade 1 (excellent image quality) to grade 5 (strongly degraded). Distinction was made between good (grades 1, 2) and poor image quality (grades 3-5). Clinical (gender, body mass index, Agatston score, heart rate and rhythm, artifacts), procedural (valve type) and technical variables (isocentricity) were related with the image quality assessment. Image quality was good in 128 (72 %) and poor in 51 (28 %) patients. By univariable analysis only valve type (BSL) and the presence of an artefact negatively affected image quality. By multivariate analysis (in which BMI was forced into the model) BSL valve (Odds 3.5, 95 % CI [1.3-9.6], p = 0.02), presence of an artifact (Odds 2.5, 95 % CI [1.2-5.4], p = 0.02) and BMI (Odds 1.1, 95 % CI [1.0-1.2], p = 0.04) were independent predictors of poor image quality. Rotational angiography with motion compensation 3D image reconstruction using a dedicated research prototype software offers good image quality for the evaluation of frame geometry after TAVI in the majority of patients. Valve type, presence of artifacts and higher BMI negatively affect image quality.
No takes yet. Share an insight, caveat, or question.
Rodríguez‐Olivares et al. (2016) conducted an observational in Aortic stenosis (n=179). Boston Sadra Lotus (BSL) valve vs. Other valve types was evaluated on Poor image quality (grades 3-5) (OR 3.5, 95% CI 1.3-9.6, p=0.02). The Boston Sadra Lotus valve, presence of artifacts, and higher body mass index were independent predictors of poor image quality in rotational angiography after TAVI.