Key result
Inadvertent LV leadless pacemaker placement via PFO linked to ischemic stroke 1 year later.
Case Report (n=1)
Preprocedural evaluation of cardiac anatomy using multiple imaging modalities is crucial to prevent inadvertent left ventricular placement of leadless pacemakers through undiagnosed septal defects.
Answers are grounded in this paper and cite their sources.
May indicate delayed stroke risk from undiagnosed PFO during Micra implantation; leaves open whether routine preprocedural imaging is warranted.
The Micra device is a leadless pacemaker implanted in the right ventricle via a femoral vein transcatheter approach. There are several indications for and advantages to using a leadless pacemaker, and the device's role in the field of cardiology will probably continue to increase. This article presents the case of a rare complication probably due to inadvertent placement of the device in the left ventricle across an undiagnosed patent foramen ovale.
No takes yet. Share an insight, caveat, or question.
Alkowatli et al. (2025) conducted a case report in Accidental left ventricular placement of a leadless pacemaker (n=1). Micra leadless pacemaker was evaluated. A 92-year-old man experienced an acute ischemic stroke one year after a Micra leadless pacemaker was inadvertently placed in the left ventricle through an undiagnosed patent foramen ovale.