Key result
Automated quantification of mitral valve tenting volume by 3D TEE distinguished moderate to severe from none to mild functional mitral regurgitation with 85% sensitivity and 71% specificity.
Why the study?
Mitral leaflet tenting contributes to functional mitral regurgitation, and a novel software tool enables automated quantification of tenting volume by 3D transesophageal echocardiography.
Does automated quantification of mitral valve tenting volume by 3D TEE accurately diagnose moderate or severe functional mitral regurgitation?
Observational (n=107)
Does automated quantification of mitral valve tenting volume by 3D TEE accurately diagnose moderate or severe functional mitral regurgitation?
Automated 3D TEE assessment of mitral valve tenting volume can help distinguish moderate/severe from none/mild functional mitral regurgitation with good sensitivity and specificity.
Answers are grounded in this paper and cite their sources.
May support automated tenting volume thresholds for functional mitral regurgitation grading; leaves open prospective validation before clinical adoption.
BACKGROUND: Tenting of the mitral leaflets is a major pathophysiological factor contributing to functional mitral regurgitation (FMR). A novel software tool allows automated quantification of the tenting volume (TnV) by 3D transesophageal echocardiography (TEE). The aims of this study are to investigate the correlations of biometric patient characteristics with the TnV and whether a threshold value for the diagnosis of a moderate or severe FMR can be calculated for the TnV. METHODS: This explorative and hypothesis-generating study analyzed the TnV of the mitral valve obtained by clinically indicated TEE. The mid-systolic, threefold calculated and averaged TnV from 80 patients with no or mild FMR and 27 patients with moderate or severe FMR was determined using the TomTec 4D MV Assessment tool. RESULTS: /m was determined for the TnV by using a receiver-operating characteristic curve. This value distinguished moderate to severe from none to mild FMR with a sensitivity of 85% and a specificity of 71%. The intra-observer variability and inter-observer variability were determined to be 0.96 and 0.85, respectively. CONCLUSIONS: Automated assessment of TnV has the potential to support the diagnostic evaluation of FMR. Further studies are needed to validate this result, detect additional factors influencing the size of the TnV, and determine further thresholds for any degree of FMR.
No takes yet. Share an insight, caveat, or question.
Hübscher et al. (2020) conducted an observational in Functional mitral regurgitation (n=107). Automated quantification of tenting volume (TnV) by 3D TEE vs. None or mild functional mitral regurgitation was evaluated on Threshold value of tenting volume for the diagnosis of moderate or severe functional mitral regurgitation. Automated quantification of mitral valve tenting volume by 3D TEE distinguished moderate to severe from none to mild functional mitral regurgitation with 85% sensitivity and 71% specificity.
Synapse has enriched 10 related papers. Consider them for comparative context: