Why the study?
Does automated quantification of mitral regurgitation using 3D-FVCD provide accurate measurements compared to PC-CMR and 2D echocardiography in patients with >moderate MR?
Does automated quantification of mitral regurgitation using 3D-FVCD provide accurate measurements compared to PC-CMR and 2D echocardiography in patients with >moderate MR?
Automated 3D-FVCD echocardiography is a feasible and accurate method for quantifying mitral regurgitation, showing better agreement with cardiac MRI than standard 2D methods.
Answers are grounded in this paper and cite their sources.
May support 3D-FVCD for MR quantification in >moderate cases; leaves open prospective validation against PC-CMR before clinical adoption.
BACKGROUND: Accurate assessment of mitral regurgitation (MR) severity is crucial for clinical decision-making and optimizing patient outcomes. Recent advances in real-time three dimensional (3D) echocardiography provide the option of real-time full volume color Doppler echocardiography (FVCD) measurements. This makes it practical to quantify MR by subtracting aortic stroke volume from the volume of mitral inflow in an automated manner. METHODS: Thirty-two patients with more than a moderate degree of MR assessed by transthoracic echocardiography (TTE) were consecutively enrolled during this study. MR volume was measured by 1) two dimensional (2D) Doppler TTE, using the proximal isovelocity surface area (PISA) and the volumetric quantification methods (VM). Then, 2) real time 3D-FVCD was subsequently obtained, and dedicated software was used to quantify the MR volume. MR volume was also measured using 3) phase contrast cardiac magnetic resonance imaging (PC-CMR). In each patient, all these measurements were obtained within the same day. Automated MR quantification was feasible in 30 of 32 patients. RESULTS: The mean regurgitant volume quantified by 2D-PISA, 2D-VM, 3D-FVCD, and PC-CMR was 72.1 ± 27.7, 79.9 ± 36.9, 69.9 ± 31.5, and 64.2 ± 30.7 mL, respectively (p = 0.304). There was an excellent correlation between the MR volume measured by PC-CMR and 3D-FVCD (r = 0.85, 95% CI 0.70-0.93, p < 0.001). Compared with PC-CMR, Bland-Altman analysis for 3D-FVCD showed a good agreement (2 standard deviations: 34.3 mL) than did 2D-PISA or 2D-VM (60.0 and 62.8 mL, respectively). CONCLUSION: Automated quantification of MR with 3D-FVCD is feasible and accurate. It is a promising tool for the real-time 3D echocardiographic assessment of patients with MR.
No takes yet. Share an insight, caveat, or question.
Son et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: