Key result
Adding 3D ICE to TEE during tricuspid TEER is linked to ~38% shorter fluoroscopy time.
Why the study?
Use of transcatheter tricuspid edge-to-edge repair is growing steadily, and the VeriSight Pro 3D ICE catheter might help increase procedural safety and improve outcomes.
Does adding 3D intracardiac echocardiography to transesophageal echocardiography guidance improve procedural outcomes and safety in patients undergoing tricuspid TEER?
Observational (n=45)
No
Does adding 3D intracardiac echocardiography to transesophageal echocardiography guidance improve procedural outcomes and safety in patients undergoing tricuspid TEER?
The addition of 3D intracardiac echocardiography to transesophageal guidance during tricuspid TEER significantly reduces fluoroscopy time and radiation exposure without compromising procedural safety.
INTRODUCTION AND OBJECTIVES: The use of transcatheter tricuspid edge-to-edge repair (TEER) is growing steadily due to its minimally invasive nature. The VeriSight Pro 3D intracardiac echocardiography (ICE) catheter might help to increase procedural safety and improve procedural outcomes. METHODS: In this prospective, single-center, observational study, we compared the procedural outcomes of tricuspid TEER guided by ICE plus transesophageal echocardiography (TEE) or by TEE alone. Lead-induced tricuspid regurgitation and combined mitral-tricuspid procedures were excluded. Primary events were those associated with procedural and early (30-day) safety. RESULTS: A total of 45 consecutive patients were included (25 guided by TEE and 20 by ICE+TEE). There were no significant differences in baseline clinical and echocardiographic features. No mortality, stroke, major bleeding, or major vascular events occurred. ICE significantly decreased the fluoroscopy time (ICE+TEE vs TEE: 16.2 [8.9-26.0] minutes vs 26 [18.2-32.0] minutes; P = .003), radiation dose (ICE+TEE vs TEE: 782 [390-1040] mGy vs 1413 [960-1705] mGy; P = .002), and dose-area product (ICE+TEE vs TEE: 5855 [2675-9328] cGy·cm CONCLUSIONS: The addition of 3D ICE guidance to TEE during tricuspid TEER is associated with lower fluoroscopy time and radiation exposure compared with TEE, without any significant disadvantages regarding procedural safety. This study represents the first national experience with the novel VeriSight Pro 3D ICE catheter.
No takes yet. Share an insight, caveat, or question.
Alperi et al. (2026) conducted an observational in Tricuspid regurgitation (n=45). 3D intracardiac echocardiography (ICE) plus transesophageal echocardiography (TEE) vs. Transesophageal echocardiography (TEE) alone was evaluated on Procedural and early (30-day) safety events. Guidance with 3D intracardiac echocardiography plus TEE during tricuspid TEER significantly reduced fluoroscopy time (16.2 vs 26.0 minutes; P=0.003) compared to TEE alone, with no safety events.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: