Key result
Cardiac output estimation using the pulse time reflection ratio from ECG and PPG strongly correlated with impedance cardiography with percent errors below the 30% clinical limit.
Why the study?
Does a novel non-invasive method using ECG and PPG accurately estimate cardiac output compared to impedance cardiography in healthy subjects during incremental exercise?
Does a novel non-invasive method using ECG and PPG accurately estimate cardiac output compared to impedance cardiography in healthy subjects during incremental exercise?
A novel non-invasive method using ECG and PPG can accurately estimate beat-to-beat cardiac output during incremental exercise in healthy subjects.
Answers are grounded in this paper and cite their sources.
May support non-invasive CO monitoring during exercise; hypothesis-generating pending validation in patients.
Cardiac output (CO) monitoring is not only essential for critically ill patients in the hospital, but also for patients at home and those undergoing cardiopulmonary exercise testing. However, CO is difficult to monitor during daily activities and exercise. In this paper, we aim at developing a novel CO estimation method that can be used under these challenging conditions. The tube model was utilized to derive a CO index, namely the pulse time reflection ratio (PTRR) from an electrocardiogram and photoplethysmogram. After calibration, the PTRR can be used to estimate beat-to-beat CO. The proposed method was verified against CO measured by impedance cardiography on 19 healthy subjects in an incremental intensity exercise test. Results showed that there were strong correlations (r) between the PTRR and reference CO in 18 subjects (mean r: 0.88, n = 245 trials). Two calibration approaches reported in the literature were applied to the proposed method and the corresponding bias +/- precisions of estimation errors were 0 +/- 1.89 L min(-1) and -0.22 +/- 2.12 L min(-1), respectively. The percent errors were 21.94% and 24.90%, smaller than the clinical acceptance limit (30%). To conclude, after calibration, this method can be used to monitor CO on healthy subjects during incremental intensity exercise.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2010) studied Healthy (n=19). Pulse time reflection ratio (PTRR) from ECG and PPG vs. Impedance cardiography was evaluated on Correlation and estimation error of cardiac output (mean r 0.88). Cardiac output estimation using the pulse time reflection ratio from ECG and PPG strongly correlated with impedance cardiography with percent errors below the 30% clinical limit.