Key result
The inflection and harmonic area ratio (IHAR) from PPG signals strongly correlated with bio-impedance cardiac output with a total percentage error of 16.2%.
Why the study?
Does the IHAR index from PPG signals accurately estimate cardiac output compared to bio-impedance in normotensive subjects?
Observational (n=16)
Does the IHAR index from PPG signals accurately estimate cardiac output compared to bio-impedance in normotensive subjects?
The IHAR index derived from PPG signals shows promise as a noninvasive, wearable method for continuous cardiac output monitoring, with acceptable error limits compared to bio-impedance.
Answers are grounded in this paper and cite their sources.
Supports IHAR for noninvasive CO monitoring in healthy subjects; leaves open clinical accuracy and utility pending validation.
Cardiac output (CO) monitoring is essential for indicating the perfusion status of the human cardiovascular system under different physiological conditions. However, it is currently limited to hospital use due to the need for either skilled operators or big, expensive measurement devices. Therefore, in this paper we devise a new CO indicator which can easily be incorporated into existing wearable devices. To this end, we propose an index, the inflection and harmonic area ratio (IHAR), from standard photoplethysmographic (PPG) signals, which can be used to continuously monitor CO. We evaluate the success of our index by testing on sixteen normotensive subjects before and after bicycle exercise. The results showed a strong intra-subject correlation between IHAR and CO(imp) measured by the bio-impedance method in fifteen subjects (mean r = 0.82, p<0.01). After least squares linear regression, the precision between CO(imp) and CO estimated from IHAR (CO(IHAR)) was 1.40 L/min. The total percentage error of the results was 16.2%, which was well below the clinical acceptance limit of 30%. The results suggest that IHAR is a promising indicator for wearable and noninvasive CO monitoring.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2009) conducted an observational in Normotensive (n=16). Inflection and harmonic area ratio (IHAR) from photoplethysmographic signals vs. Bio-impedance method was evaluated on Intra-subject correlation between IHAR and cardiac output measured by bio-impedance (mean r = 0.82, p=<0.01). The inflection and harmonic area ratio (IHAR) from PPG signals strongly correlated with bio-impedance cardiac output with a total percentage error of 16.2%.