April 27, 2010Blood Pressure Monitoring254 citations

Evaluation of the Mobil-O-Graph new generation ABPM device using the ESH criteria

View Full Paper
PFPascal ML. FranssenBIBen P.M. Imholz

Key Result

The Mobil-O-Graph ambulatory blood pressure monitor passed all ESH validation phases, with systolic pressure differences within 15, 10, and 5 mmHg in 98%, 94%, and 71% of measures, respectively.

Study Design

Type

Observational (n=33)

Structured PICO

Does the Mobil-O-Graph new generation ABPM device accurately measure blood pressure according to the European Society of Hypertension criteria?

P
Population
33 individuals undergoing ambulatory blood pressure monitor validation according to European Society of Hypertension criteria.
E
Exposure
Mobil-O-Graph new generation 24/48 h ambulatory blood pressure monitor
C
Comparator
Reference blood pressure measurements according to European Society of Hypertension criteria
O
Outcome
Accuracy of systolic and diastolic blood pressure measurements within 15, 10, and 5 mmHg limits according to ESH criteriasurrogate

The Mobil-O-Graph new generation ambulatory blood pressure monitor meets the European Society of Hypertension accuracy criteria and is recommended for clinical practice.

Abstract

We report on the validation of the new generation Mobil-O-Graph 24/48 h ambulatory blood pressure monitor according to the criteria of the European Society of Hypertension. In 15 individuals participating in phase I for systolic pressure, all 45 measures differed less than 15 mmHg, 43 and 33 out of 45 differed less than 10 and 5 mmHg. As for diastolic pressures even better scores were reached when the device passed the EHS score. In phase II, data were collected in an additional 18 individuals leaving a total of 33 individuals and 99 measures. The phase counts the achieved percentages of two or three measures per individual within 15, 10 and 5 mmHg limits. Systolic pressures exceeded the required 95, 80 and 65% for 15, 10 and 5 mmHg differences with values of 98, 94 and 71%, respectively. As again for diastolic pressure the values were even better, the device passed phase II also. Thus, all phases of the European Society of Hypertension procedure were passed and the results of this study can recommend the use of the Mobil-O-Graph new generation ambulatory blood pressure monitor device in clinical practice.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Franssen et al. (2010) reported an observational. Mobil-O-Graph new generation ABPM device vs. ESH criteria reference was evaluated on Percentage of measures within 15, 10 and 5 mmHg limits according to ESH criteria. The Mobil-O-Graph ambulatory blood pressure monitor passed all ESH validation phases, with systolic pressure differences within 15, 10, and 5 mmHg in 98%, 94%, and 71% of measures, respectively.

synapsesocial.com/papers/6a3301335fe356619419ec40https://doi.org/10.1097/mbp.0b013e328339be38
Ask AI
Helpful
Bookmark
Share
View Full Paper