Key result
ACURATE neo2 linked to more symmetrical implantation depth vs Portico and Evolut in complex landing zones.
Why the study?
Increased aortic angulation can affect self-expandable TAVI implantation depth, but the behavior of ACURATE neo2 across patient-specific aortic landing zone geometry was uncharacterized.
Does the ACURATE neo2 self-expanding valve lead to a more predictable and symmetrical implantation depth compared with Portico/Evolut valves in severe aortic stenosis patients, based on aortic landing zone geometry?
Cohort (n=207)
No
Does the ACURATE neo2 self-expanding valve lead to a more predictable and symmetrical implantation depth compared with Portico/Evolut valves in severe aortic stenosis patients, based on aortic landing zone geometry?
The ACURATE neo2 valve's unique top-down deployment and stabilization arches result in a symmetrical implantation depth that is minimally affected by aortic landing zone angulation and curvature, unlike the Portico and Evolut self-expandable valves.
Answers are grounded in this paper and cite their sources.
ACURATE neo2 may yield more predictable depth in angulated aortas; hypothesis-generating and should not yet change TAVI device selection.
Background During transcatheter aortic valve implantation (TAVI), increased aortic angulation can affect the final implantation depth of self-expandable (SE) devices due to the interaction between the high stent frame and the targeted aortic landing zone (LZ). We herein sought to investigate the behaviour of the ACURATE neo 2, a SE device with a unique release mechanism, in relation to patient-specific angulation and curvature of the aortic LZ. Methods The mismatch between the intended and the final implantation depth (∆H) was compared between patients treated with ACURATE neo2 (Acurate, n=106) and Evolut/Portico (n=101) SE devices. To do so, curvature ( κ LZ,tot ) and angulation ( α LZ,Dist ) were calculated based on the three-dimensional aortic LZ centerline available from pre-TAVI CT. Results The Acurate and Evolut/Portico groups showed a negligible difference (p=0.09) for ∆H averaged between non-coronary (NCC) and left coronary cusp (LCC). However, when splitting both ∆H NCC and ∆H LCC values into two subgroups based on κ LZ,tot and α LZ,Dist median values, ∆H significantly increased on LCC compared with NCC in Evolut/Portico patients with high LZ curvature (κ LZ,tot >0.123/mm, p=0.016) and high LZ distal angulation ( α LZ,Dist >28.5°, p=0.012). No statistically significant differences arose within the Acurate group. Conclusions Among SE devices, the ACURATE neo 2 was the least affected by the curvature and angulation of the LZ anatomy, leading to a more predictable and symmetrical implantation depth. The clinical impact of this finding on TAVI outcomes in patients with an angulated aortic LZ warrants further investigation in larger studies.
No takes yet. Share an insight, caveat, or question.
Sturla et al. (2025) conducted a cohort in Severe aortic stenosis (n=207). ACURATE neo2 vs. Evolut/Portico self-expandable devices was evaluated on Mismatch between the intended and the final implantation depth (∆H averaged between non-coronary and left coronary cusp) (p=0.09). ACURATE neo2 achieved a symmetrical implantation depth closely matching the intended target and was significantly less affected by aortic landing zone curvature and angulation compared to Portico and Evolut devices.
Synapse has enriched 10 related papers. Consider them for comparative context: