Answers are grounded in this paper and cite their sources.
Study developed a device for detecting CEA in serum, suggesting a novel tool for point-of-care testing.
This work initially investigated potential changes in the surface chemistry of cellulose induced by oxygen plasma treatment through the application of complementary optical surface characterization techniques. Subsequently, a simplified paper-based metalloimmunoassay for carcinoembryonic antigen (CEA) detection was developed, employing plasma-treated paper disks as the functional substrate. In-house synthesized CdS quantum dots (QDs) were directly bio-conjugated to reporter anti-CEA antibody and were utilized as fluorescent labels using simple instrumentation for detection. The plasma-treated paper demonstrated enhanced analytical sensitivity, attributed to the increased exposure of cellulose fibers following treatment. The proposed assay achieved a limit of detection of 0.09 ng mL⁻¹ for CEA, significantly below the clinical cut-off value of 5 ng mL⁻¹. The recoveries in serum samples were in the range 103–120%, with relative standard deviations ranging from 6.2% to 12.4%. The paper-based metalloimmunoassay represents a field-deployable sensing tool with potential for point-of-care applications, particularly in resource-limited settings.
No takes yet. Share an insight, caveat, or question.
Mermiga et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: