Key result
TEER causes elevated E/e' ratio, increased radial strain, and mild non-progressive mitral stenosis.
Why the study?
Mitral valve regurgitation is common in canines, but no detailed information has been published on the hemodynamic effect of TEER on cardiac function.
Does Transcatheter Edge-to-Edge Repair (TEER) alter cardiac function in healthy dogs?
Does Transcatheter Edge-to-Edge Repair (TEER) alter cardiac function in healthy dogs?
Answers are grounded in this paper and cite their sources.
TEER can be safely implanted in a small canine model using a hybrid approach, providing a minimally invasive platform for mitral valve repair, though it induces expected functional regurgitation and mild stenosis when applied to normal valves.
Mitral valve regurgitation is a common canine heart disease. Transcatheter Edge-to-Edge Repair (TEER) is a transcatheter, edge-to-edge mitral repair device that uses a hybrid approach. No detailed information has been published on the hemodynamic effect of TEER on cardiac function. The aim of this report is to provide a longitudinal observation of the cardiac functional changes observed after TEER implantation in normal dogs using traditional, two-dimensional speckle tracking, and color M-mode echocardiographic methods. In the current report, TEER was implanted into two healthy dogs under general anesthesia. An echocardiographic examination was performed at baseline and weekly postoperative follow-ups were conducted until the fourth week. Successful TEER implantation was achieved with a short operation time (98 and 63 min) in the two dogs. Functional mitral valve regurgitation, elevated E/e' ratio, elevated radial strain, and stable intraventricular pressure gradients (IVPG) were observed after the operation in the dogs. Mild non progressive mitral valve stenosis was observed in both dogs. TEER is a minimally invasive method for mitral valve surgery that necessitates more clinical trials. With longitudinal observation of heart function using novel approaches, better outcomes will be expected.
No takes yet. Share an insight, caveat, or question.
Sasaki et al. (2021) studied Healthy dogs (n=2). Transcatheter Edge-to-Edge Repair (TEER) was evaluated on Echocardiographic changes in cardiac function. Transcatheter Edge-to-Edge Repair (TEER) was successfully implanted with operation times of 98 and 63 minutes, resulting in functional mitral valve regurgitation, elevated E/e' ratio, elevated radial strain, and mild non-progressive mitral stenosis.