Key result
Subcutaneous implantable cardioverter-defibrillator implantation and defibrillation threshold testing caused the aggravation and displacement of an existing, incompletely healed clavicle fracture.
Case Report (n=1)
No
Defibrillation threshold testing during S-ICD implantation can cause displacement of incompletely healed clavicle fractures due to vigorous chest wall muscle contraction.
Answers are grounded in this paper and cite their sources.
Warrants caution implanting S-ICD near incompletely healed clavicle fractures; leaves open optimal device selection in young ARVC patients.
An 18 year old male with an incompletely healed clavicle fracture presented with unexplained syncope. Subsequent investigations were consistent with a diagnosis of arrhythmogenic right ventricular cardiomyopathy (ARVC). A subcutaneous implantable cardioverter-defibrillator (S-ICD) was successfully implanted and defibrillation threshold (DFT) testing performed as per standard protocol. Shortly following the procedure, the patient complained of pain and swelling over the left clavicle. A radiograph revealed aggravation and displacement of the underlying clavicle fracture. Surgical reduction and internal fixation was performed one week later.
No takes yet. Share an insight, caveat, or question.
Ismail et al. (2020) conducted a case report in Arrhythmogenic right ventricular cardiomyopathy (ARVC) (n=1). Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation with defibrillation threshold testing was evaluated on Displacement of existing clavicle fracture. Subcutaneous implantable cardioverter-defibrillator implantation and defibrillation threshold testing caused the aggravation and displacement of an existing, incompletely healed clavicle fracture.