Key result
A 50-year-old male developed Twiddler's syndrome 4 weeks after subcutaneous-ICD implantation, presenting with an inappropriate shock and lead dislodgement requiring revision.
Why the study?
Twiddler syndrome is reported in transvenous devices, but there is a paucity of data on subcutaneous devices.
Case Report (n=1)
Twiddler's syndrome, a rare complication involving device displacement, can occur with subcutaneous ICDs and present with inappropriate shocks and abnormal lead impedance.
Answers are grounded in this paper and cite their sources.
Twiddler's syndrome warrants consideration in S-ICD patients with inappropriate shocks; this case leaves open its incidence and prevention strategies.
Twiddler’s syndrome is a rare complication where a pacemaker or implantable cardioverter-defibrillator (ICD) is displaced with or without patient manipulation of their device. There are reports on transvenous devices but a paucity of data on subcutaneous devices. A 50-year-old male with hypertrophic cardiomyopathy and non-sustained ventricular tachycardia underwent subcutaneous-ICD (S-ICD) implantation for primary prevention. Remote device interrogation 4 weeks after device placement reported a shock due to “ventricular fibrillation”. It also showed abnormal lead impedance. Chest X-ray showed lead was dislodged and coiled around the pulse generator. Patient underwent lead revision and device replacement without further complications.
No takes yet. Share an insight, caveat, or question.
Jin et al. (2020) conducted a case report in Hypertrophic cardiomyopathy and non-sustained ventricular tachycardia (n=1). Subcutaneous-ICD (S-ICD) implantation was evaluated on Twiddler's syndrome (lead dislodgement). A 50-year-old male developed Twiddler's syndrome 4 weeks after subcutaneous-ICD implantation, presenting with an inappropriate shock and lead dislodgement requiring revision.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: