Key result
Twiddler syndrome causes S-ICD lead failure requiring extraction and submuscular reimplantation.
Case Report (n=1)
No
Twiddler syndrome in S-ICD patients can cause lead failure due to mechanical stress, which may be prevented by submuscular reimplantation.
Answers are grounded in this paper and cite their sources.
May prompt vigilance for Twiddler syndrome in S-ICD malfunction; submuscular reimplantation leaves optimal prevention open for validation.
We present a case of Twiddler syndrome in a patient with a subcutaneous implantable cardioverter defibrillator (S-ICD). The patient presented herself to the outpatient clinic with pain in the left chest. Chest x-ray confirmed Twiddler syndrome and ICD read-out revealed lead failure resulting in absent heart rhythm sensing in one vector. The lead and pulse generator were extracted and a new S-ICD system was reimplanted submuscular underneath the serratus anterior muscle to prevent reoccurrence. Lead investigation revealed an insulation defect caused by excessive mechanical stress.
No takes yet. Share an insight, caveat, or question.
Kooiman et al. (2015) conducted a case report in Twiddler syndrome with subcutaneous implantable cardioverter defibrillator (S-ICD) (n=1). Submuscular reimplantation of S-ICD was evaluated. Twiddler syndrome in a patient with an S-ICD caused lead failure and absent sensing, requiring extraction and submuscular reimplantation of a new system to prevent recurrence.
Synapse has enriched 10 related papers. Consider them for comparative context: