Twiddler's syndrome can be detected prior to serious implantable cardioverter-defibrillator malfunction based on a patient's complaint of generator rotation in the subcutaneous pocket.
Case Report (n=1)
Patient complaints of ICD generator rotation can facilitate early detection of Twiddler's syndrome before serious device malfunction occurs.
Answers are grounded in this paper and cite their sources.
Patient-reported ICD generator rotation warrants prompt evaluation; leaves open whether proactive monitoring prevents Twiddler's syndrome complications.
Twiddler's syndrome presents as a rare implantable‐cardioverter defibrillator (ICD) malfunction that is potentially fatal. The following case is that of a 79‐year‐old woman, who was implanted with an ICD and experienced Twiddler's syndrome. Five months after the implantation, the patient complained that the generator sometimes rotated in the subcutaneous pocket. Fluoroscopy showed that the generator had rotated and the leads were twisted, which is compatible with Twiddler's syndrome. Although all the leads were functioning normally, a procedure to repair the twisted leads was performed. Twiddler's syndrome is usually detected due to a serious device malfunction; however, this case study suggests that it can be detected prior to ICD malfunction by the patient's complaint.
No takes yet. Share an insight, caveat, or question.
Konishi et al. (2012) conducted a case report in Twiddler's syndrome (n=1). Implantable cardioverter-defibrillator (ICD) was evaluated on Detection of Twiddler's syndrome. Twiddler's syndrome can be detected prior to serious implantable cardioverter-defibrillator malfunction based on a patient's complaint of generator rotation in the subcutaneous pocket.
Synapse has enriched 10 related papers. Consider them for comparative context: