This case report highlights that patients with internal automatic cardioverter-defibrillators are also susceptible to Twiddler's syndrome, previously described mainly in pacemaker patients.
Answers are grounded in this paper and cite their sources.
Raises awareness for Twiddler's syndrome in ICD malfunction; extends pacemaker reports but leaves incidence and risks open.
Twiddler's syndrome, characterized by dislodgment of pacemaker leads due to twisting of pulse generators within the subcutaneous pocket with subsequent retraction of leads and loss of pacing function, has been well described in patients with permanent pacemakers implanted for bradyarrhythmias. The case presented here is the first report of a patient with an internal automatic cardioverter-defibrillator with lead dislodgment due to pulse generator rotation. This case exemplifies a new subset of patients prone to the Twiddler's syndrome.