In a 48-year-old man with a DDD pacemaker, the MARKER CHANNEL confirmed that alternating QRS morphologies were due to atrial lead dislodgement despite normal chest x-rays.
Case Report (n=1)
The MARKER CHANNEL of a DDD pulse generator can aid in diagnosing atrial lead dislodgement when chest x-rays appear normal.
Answers are grounded in this paper and cite their sources.
Marker channel interrogation may diagnose atrial lead dislodgement missed on x-ray; supports targeted use of device diagnostics in troubleshooting.
A 48-year-old man with previous aortic valve surgery and aortic root repair had a DDD pacemaker inserted (using transvenous leads) for the treatment of complete heart block. An atrial J active fixation electrode was used. Four weeks following implantation the patient returned with an unusual electrocardiographic rhythm demonstrating two separate QRS morphologies. Both PA and lateral chest x-ray failed to demonstrate a change in lead position. The MARKER CHANNEL of the DDD pulse generator confirmed that the alternating QRS morphologies were due to atrial lead dislodgement. Although uncommon, displacement of atrial active fixation leads may occur and lead to unusual electrocardiographic rhythms. Use of a MARKER CHANNEL may aid in the diagnosis.
No takes yet. Share an insight, caveat, or question.
Greenspon et al. (1986) conducted a case report in Complete heart block (n=1). DDD pacemaker insertion with atrial J active fixation electrode was evaluated on Atrial lead dislodgement. In a 48-year-old man with a DDD pacemaker, the MARKER CHANNEL confirmed that alternating QRS morphologies were due to atrial lead dislodgement despite normal chest x-rays.
Synapse has enriched 10 related papers. Consider them for comparative context: