Key result
Implantation of an atrial J electrode in a patient with prior right atrial appendage removal resulted in lead dislodgement, suggesting screw-in electrodes offer superior stability in such cases.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
In patients with prior right atrial appendage removal, an atrial J electrode is unstable and prone to dislodgement, making a screw-in atrial electrode the preferred choice.
A DDD pacemaker utilizing an atrial J electrode was implanted in a patient whose right atrial appendage had been removed 7 years prior during aortic valve replacement. Alternating QRS morphologies produced by pacemaker stimulus artifacts occurred as a consequence of atrial lead dislodgement. The screw-in atrial electrode has superior stability in patients with right atrial appendage removal; stability of the atrial J electrode cannot be obtained in such cases.
No takes yet. Share an insight, caveat, or question.
Stark et al. (1985) conducted a case report in DDD pacemaker system malfunction (n=1). Atrial J electrode vs. Screw-in atrial electrode was evaluated on Atrial lead dislodgement and alternating QRS morphologies. Implantation of an atrial J electrode in a patient with prior right atrial appendage removal resulted in lead dislodgement, suggesting screw-in electrodes offer superior stability in such cases.