Key result
3D wideband LGE MRI suppresses ICD artifacts to enable volumetric scar delineation vs conventional narrowband scans.
Answers are grounded in this paper and cite their sources.
May improve VT target delineation amid ICD artifacts; leaves open whether outcomes improve in prospective trials.
PURPOSE Cardiac stereotactic body radiotherapy (SBRT) offers a noninvasive alternative to the conventional radiofrequency (RF) catheter ablation for treating refractory ventricular tachycardia (VT) in patients who are poor candidates for ablation procedures. However, accurate target delineation remains challenging due to limited scar visualization on CT and implanted cardioverter-defibrillator (ICD)-related MRI artifacts. We developed and evaluated a 3D wideband (WB) late gadolinium-enhanced (LGE) MRI protocol for artifact suppression, geometric accuracy, and scar delineation in SBRT planning. METHODS AND MATERIALS The 3D WB LGE sequence employs a 6-kHz wideband hyperbolic secant inversion pulse to suppress ICD-induced hyperintense artifacts followed by a 3D gradient-echo readout. Both electrocardiogram (ECG) and respiratory navigator gating were employed to manage the cardiac and respiratory motion, respectively. Protocol optimization was performed in phantoms and four healthy volunteers with ICDs positioned near the chest. Geometric distortion was measured using MagPhan and American College of Radiology (ACR) phantoms. Clinical feasibility was tested in a 73-year-old patient undergoing cardiac SBRT. The 3D LGE MRI was registered with the planning CT and compared with electroanatomic mapping (EAM) using the AHA 17-segment model. RESULTS Phantom analysis showed minimal sequence-related distortion, well within the clinical tolerance. ICD-induced distortion was negligible when the device was placed 10 cm away from the target. The WB inversion pulse successfully suppressed ICD-related hyperintense artifacts in phantoms and healthy volunteers compared to conventional narrowband (NB) LGE MRI. In the VT patient, 3D WB LGE MRI enabled volumetric scar delineation with good correlation between gadolinium enhancement and EAM-defined targets. Acquisition time for the 3D WB LGE MRI was optimized to be approximately 5 minutes. CONCLUSION 3D WB LGE MRI is feasible for cardiac SBRT simulation, achieving reliable ICD artifact suppression and minimal distortion while providing good scar visualization non-invasively. Its integration into clinical workflows may improve VT target delineation accuracy in patients with implanted devices.
No takes yet. Share an insight, caveat, or question.
Dong et al. (2026) studied Refractory ventricular tachycardia (n=5). 3D wideband (WB) late gadolinium-enhanced (LGE) MRI vs. Conventional narrowband (NB) LGE MRI was evaluated on Artifact suppression, geometric accuracy, and scar delineation. A 3D wideband LGE MRI protocol successfully suppressed ICD-related hyperintense artifacts and enabled volumetric scar delineation with minimal distortion compared to conventional narrowband LGE MRI.
Synapse has enriched 10 related papers. Consider them for comparative context: