Examines normalization quality assurance in PET imaging, indicating implications for clinical deployment.
Quantitative PET imaging depends critically on the accuracy of system normalization and uniformity calibration. We report a manufacturing defect in a certified RadQual ⁶⁸Ge PET cylinder phantom (BMCY68-2030A-03), a standard reference used for normalization and daily quality control in Siemens Biograph Vision scanners. A localized hot spot was identified within the phantom, producing integral uniformity greater than 8.5% (acceptance limit ≤5%). The nonuniformity was intrinsic to the phantom material and rendered it unsuitable for quantitative use. This finding underscores the importance of independent acceptance testing of long-lived ⁶⁸Ge standards before clinical deployment to prevent calibration bias and ensure reliable SUV quantification.