Key result
Immediate back-to-back rest/stress flurpiridaz PET with residual correction closely matches separate-scan MBF reference values.
Why the study?
18F-flurpiridaz results in high residual tracer activity during same-day rest/stress scanning, prompting a technical evaluation of immediate back-to-back rest/stress PET with software correction for residual activity effects.
Does software correction for residual activity enable accurate immediate back-to-back rest/stress 18F-flurpiridaz PET scanning?
Does software correction for residual activity enable accurate immediate back-to-back rest/stress 18F-flurpiridaz PET scanning?
p-value: p=<.001
Software correction for residual tracer activity enables accurate immediate back-to-back rest and stress 18F-flurpiridaz PET scanning without delay.
May enable same-day rest-stress flurpiridaz PET; leaves open prospective validation before clinical use.
BACKGROUND: F enables flurpiridaz to be made and shipped regionally; however, it also results in high levels of residual tracer activity for same-day rest/stress scanning. This study provides a technical evaluation of immediate back-to-back rest/stress flurpiridaz positron emission tomography (PET) with no delay, using a commercially available software medical device (mfiVerse™, MultiFunctional Imaging LLC) to correct for residual activity effects. METHODS AND RESULTS: Retrospectively obtained separate-scan dynamic rest and pharmacologic stress flurpiridaz images in 11 subjects were digitally shifted and combined to model a 15 minute rest/stress procedure with rest scan from 0 to 8 minutes and stress scan from 8 to 15 minutes, providing a paired-sample study design. The modeling of residual tracer effects included both distributed activity and stress effects upon residual arterial input. Residual activity biased mean stress pixel values and myocardial blood flow (MBF) by +31% and +64%, respectively, as compared with the reference stress images without residual activity. These biases were each reduced to less than 1% after residual activity correction. Linear regression analysis found that corrected stress pixel values (r = .998) and MBF (r = .994) closely matched reference stress values, both being equivalent to separate-scan values to less than half of test-retest variability (P < .001). Summed stress scores averaged 2.5 points lower for uncorrected images versus the reference images, and averaged .4 points lower after correction. CONCLUSIONS: Within the study limitations, these results demonstrate that rest and stress flurpiridaz PET images can be obtained immediately back-to-back without rest-stress delay when accurate correction for residual activity is performed.
No takes yet. Share an insight, caveat, or question.
Kadrmas et al. (2025) studied this question. Immediate back-to-back rest/stress flurpiridaz PET with residual activity correction vs. Reference stress images without residual activity was evaluated on Equivalence of corrected stress pixel values and myocardial blood flow to reference separate-scan values (p=<.001). Immediate back-to-back rest/stress flurpiridaz PET with residual activity correction closely matched reference separate-scan stress pixel values (r=0.998) and myocardial blood flow (r=0.994, P<0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: