Key result
Optimal kissing balloon inflation in a coronary bifurcation model can be achieved with minimal balloon overlapping and proximal dilatation by an optimally sized balloon.
Why the study?
Does minimal balloon overlapping (MBO) with proximal dilatation improve stent configuration compared to operator discretion or long balloon overlapping in a coronary bifurcation model?
Does minimal balloon overlapping (MBO) with proximal dilatation improve stent configuration compared to operator discretion or long balloon overlapping in a coronary bifurcation model?
Minimal balloon overlapping with proximal dilatation by an optimally sized balloon achieves optimal kissing balloon inflation in coronary bifurcation models.
AIMS: To define the optimal kissing balloon inflation (KBI) after single-stent deployment in a coronary bifurcation model. METHODS AND RESULTS: We deployed stents in main vessels (MV) followed by KBI in various conditions and compared the stent configurations. A) KBI at the operator's discretion vs. under the guidelines of minimal balloon overlapping (MBO). Various stent configurations were observed after the former option, whereas similar maximal dilation points were observed under the MBO guidelines. B) Long balloon overlapping (LBO) vs. MBO with proximal MV dilated by a large balloon. The proximal MV was dilated to an ideal round shape with MBO versus an oval shape with LBO. C) Two-link vs. 3-link stents. Although the 2-link stent was advantageous to open the side branch, it incurred a risk of overdilatation of the proximal struts, whereas the 3-link stent preserved its structure. Computed simulations of coronary flow were analysed in the following left main coronary models: circle with a diameter of 4 and 5.5 mm, ellipse with longitudinal direction and tilt position. They revealed that the overdilated side was exposed to low shear stress regardless of its shape. CONCLUSIONS: Optimal KBI can be achieved with MBO and proximal dilatation by an optimally sized balloon.
No takes yet. Share an insight, caveat, or question.
Murasato et al. (2014) studied Coronary bifurcation. Kissing balloon inflation (KBI) under minimal balloon overlapping (MBO) guidelines vs. KBI at operator's discretion and long balloon overlapping (LBO) was evaluated on Stent configuration and computed simulations of coronary flow. Optimal kissing balloon inflation in a coronary bifurcation model can be achieved with minimal balloon overlapping and proximal dilatation by an optimally sized balloon.