Key result
Smartwatch heart rate detection achieves 100% success in LVAD patients, but SpO2 acquisition largely fails.
Why the study?
Consumer wearables are increasingly used for self-monitoring, but their performance in patients with continuous-flow LVADs is uncertain.
Does consumer smartwatch-based biosignal acquisition accurately measure heart rate, SpO2, and ECG in patients with advanced heart failure with and without LVAD support compared to clinical references?
Observational (n=30)
No
Does consumer smartwatch-based biosignal acquisition accurately measure heart rate, SpO2, and ECG in patients with advanced heart failure with and without LVAD support compared to clinical references?
Answers are grounded in this paper and cite their sources.
Consumer smartwatches can reliably measure heart rate and frequently provide interpretable single-lead ECGs in patients with continuous-flow LVADs, but SpO2 acquisition remains unreliable.
Background Consumer wearables are increasingly used for self-monitoring, but their performance in patients with continuous-flow LVADs is uncertain. We evaluated feasibility and accuracy of smartwatch-based biosignal acquisition in advanced heart failure (HF) with and without LVAD support. Methods In a prospective, single-center pilot, 30 patients were enrolled (15 AHA/ACC stage D HF, 15 continuous-flow LVAD). Each underwent parallel measurements with a consumer smartwatch and clinical references: cuff blood pressure, fingertip SpO2, ECG-derived heart rate (HR), and a 12-lead ECG. Following skin antisepsis, the device was secured to the wrist; HR recording excluded an initial 20-s calibration. Simultaneous SpO2 and three 30-s single-lead ECGs were acquired. Up to three attempts per modality were allowed; failures after three attempts were deemed unsuccessful. Agreement was assessed by Bland–Altman analysis. Results Thirty patients were included: heart failure (n = 15; age 65 ± 5 years; LVEF 25 ± 2%) and LVAD (n = 15; age 62 ± 2 years; LVEF 17 ± 2%; flow 4.5 ± 0.2 L/min). In the HF cohort, smartwatch data acquisition demonstrated high success for HR, SpO2, and single-lead ECG (overall success 93%). QRS width on the smartwatch lead deviated from standard lead I by 16±4% with 33% of patients within a 10% margin (95% CI, 15–58%). In LVAD patients, HR detection was feasible in 100% and accurate in 80% (≤5% deviation) and 93% (≤10% deviation), independent of aortic valve opening. Bland–Altman bias was −0.1 bpm (SD 5.7 bpm; limits of agreement [LoA] −11.3 to 11.1 bpm). Single-lead ECGs were interpretable in 73%. All non-interpretable traces exhibited QRS amplitudes <0.3 mV relative to baseline noise, suggesting a practical amplitude cut-off. QRS width deviation was 12±4%, with 53% (95% CI, 30–75%) within a 10% margin; bias was −5.3 ms (SD 16 ms; LoA −36.7 to 26.1 ms). SpO2 monitoring largely failed (20% success) without identifiable predictors of success (hair density, blood pressure, LVAD flow, and pulse index not associated). Bland–Altman bias was −1.7% (SD 2.1%; LoA −5.8% to 2.4%). Conclusions In advanced HF without LVADs, smartwatch acquisition of HR, SpO2, and ECG signals was largely feasible. In continuous-flow LVAD patients, smartwatches provided reliable HR and frequently interpretable single-lead ECGs, whereas SpO2 acquisition remained unreliable. These findings delineate the potential and current limitations of wearable-based telemonitoring in LVAD populations and underscore the need for algorithmic and hardware optimization to improve oximetry and low-amplitude ECG detection.
No takes yet. Share an insight, caveat, or question.
Heil et al. (2026) conducted an observational in Advanced heart failure with and without LVAD support (n=30). Consumer smartwatch vs. Clinical reference devices was evaluated on Feasibility and accuracy of smartwatch-based biosignal acquisition (HR, SpO2, and single-lead ECG). In patients with continuous-flow LVADs, smartwatch heart rate detection was 100% feasible and accurate (bias -0.1 bpm), but SpO2 acquisition largely failed with only 20% success.