Key result
In a patient-specific computational model with IVCD and lateral LV scar, LOT-CRT achieved the shortest left ventricular activation time (108 ms) compared to BiVP (129 ms) and nsLBBP (167 ms).
Why the study?
While LBBAP may correct proximal left bundle branch block, its electrical effects in the presence of intraventricular conduction disease and myocardial scar remain unclear, prompting the creation of a computational framework to determine optimal CRT strategies.
Does patient-specific computational modeling of different CRT strategies identify the optimal pacing approach for a patient with IVCD and LV scar?
Does patient-specific computational modeling of different CRT strategies identify the optimal pacing approach for a patient with IVCD and LV scar?
Patient-specific computational modeling can simulate electrical responses to various CRT strategies, demonstrating that LOT-CRT achieved the highest degree of LV synchrony in a patient with IVCD and lateral LV scar.
Background In patients with dyssynchronous heart failure, cardiac resynchronization therapy (CRT) is the preferred treatment. As a more physiological alternative to biventricular pacing (BiVP), left bundle branch area pacing (LBBAP) and LBBAP optimized CRT (LOT-CRT) are gaining traction. While LBBAP may correct a pure proximal left bundle branch block, its electrical effects in the presence of intraventricular conduction disease (IVCD) and myocardial scar remain unclear. The purpose of this study is to create a computational modelling framework which can determine the optimal CRT strategy for a specific patient. Methods To demonstrate the modelling workflow, the framework was applied to a single patient-derived dataset. A biventricular geometry was reconstructed from computed tomography, scar regions were segmented from delayed enhancement cardiac MRI and modelled as a mixture of non-conductive and slow-conducting tissue. Fibre architecture and Purkinje network were generated using a rule-based approach. Electrical activation was simulated using an Eikonal model. Body-surface ECGs were computed using lead-field theory. Model parameters, including Purkinje topology, conduction velocities, the fraction of non-conductive scar, and the interventricular conduction delay, were optimized using a Bayesian approach. The calibrated model was used as baseline and compared with simulations of three pacing strategies: conventional BiVP; non-selective left bundle branch pacing (nsLBBP), achieved by pacing the LBB and surrounding myocardium; and LOT-CRT, achieved by simultaneously pacing the nsLBBP site and the LV free wall. Results The RMSE between simulated and measured ECGs was ~0.1 mV, with morphologies and QRS duration matching the measured ECGs, indicating a realistic reconstruction of the intrinsic activation pattern. Notably, optimization resulted in a significantly reduced Purkinje conduction velocity, suggesting underlying diffuse conduction disease. Figure 1 shows the intracardiac activation sequences and corresponding simulated ECGs for baseline (intrinsic), BiVP, nsLBBP, and LOT-CRT. The total left ventricular activation time was 150 ms during intrinsic activation, 129 ms during BiVP, 167 ms during nsLBBP, and 108 ms during LOT-CRT. Simulated QRS durations were 134 ms during intrinsic activation, 106 ms during BiVP, 134 ms during nsLBBP, and 134 ms during LOT-CRT. Conclusion Patient-specific computational modelling in a patient with IVCD, and lateral LV myocardial scar demonstrated that, for this patient, nsLBBP alone resulted in less synchronous electrical activation than BiVP, while LOT-CRT achieved the highest degree of synchrony of the LV. These results demonstrate patient-specific models may help identify the optimal CRT strategy.
No takes yet. Share an insight, caveat, or question.
Gerrits et al. (2026) studied Dyssynchronous heart failure (n=1). LBBAP optimized CRT (LOT-CRT) and non-selective left bundle branch pacing (nsLBBP) vs. Conventional biventricular pacing (BiVP) and intrinsic baseline was evaluated on Total left ventricular activation time. In a patient-specific computational model with IVCD and lateral LV scar, LOT-CRT achieved the shortest left ventricular activation time (108 ms) compared to BiVP (129 ms) and nsLBBP (167 ms).