Key result
Automated computer-based analysis of continuous ECG telemetry for AF detection in stroke patients had a specificity of 94.6% and sensitivity of 78.4% compared to manual analysis.
Why the study?
Manual analysis of continuous ECG monitoring to detect AF on stroke units requires considerable resources, while automated computer-based analysis of RR intervals may simplify detection, prompting a head-to-head comparison of both methods.
Does automated computer-based analysis of continuous telemetry accurately detect atrial fibrillation compared to manual analysis in stroke patients?
Observational (n=216)
Blinded
Does automated computer-based analysis of continuous telemetry accurately detect atrial fibrillation compared to manual analysis in stroke patients?
Automated computer-based AF detection in stroke units has high specificity but low sensitivity, indicating it should be used as an adjunct to manual ECG analysis rather than a replacement.
Automated AF detection shows high specificity but modest sensitivity in stroke telemetry; hypothesis-generating for adjunctive use pending validation.
BACKGROUND: Detection of atrial fibrillation (AF) is one of the primary diagnostic goals for patients on a stroke unit. Physician-based manual analysis of continuous ECG monitoring is regarded as the gold standard for AF detection but requires considerable resources. Recently, automated computer-based analysis of RR intervals was established to simplify AF detection. The present prospective study analyzes both methods head to head regarding AF detection specificity, sensitivity, and overall effectiveness. METHODS: Consecutive stroke patients without history of AF or proof of AF in the admission ECG were enrolled over the period of 7 months. All patients received continuous ECG telemetry during the complete stay on the stroke unit. All ECGs underwent automated analysis by a commercially available program. Blinded to these results, all ECG tracings were also assessed manually. Sensitivity, specificity, time consumption, costs per day, and cost-effectiveness were compared. RESULTS: 216 consecutive patients were enrolled (70.7 ± 14.1 years, 56% male) and 555 analysis days compared. AF was detected by manual ECG analysis on 37 days (6.7%) and automatically on 57 days (10.3%). Specificity of the automated algorithm was 94.6% and sensitivity 78.4% (28 [5.0%] false positive and 8 [1.4%] false negative). Patients with AF were older and had more often arterial hypertension, higher NIHSS at admission, more often left atrial dilatation, and a higher CHA2DS2-VASc score. Automation significantly reduced human resources but was more expensive compared to manual analysis alone. CONCLUSION: Automatic AF detection is highly specific, but sensitivity is relatively low. Results of this study suggest that automated computer-based AF detection should be rather complementary to manual ECG analysis than replacing it.
No takes yet. Share an insight, caveat, or question.
Rogalewski et al. (2020) conducted an observational in Stroke (n=216). Automated computer-based analysis of continuous ECG telemetry vs. Physician-based manual analysis of continuous ECG monitoring was evaluated on Atrial fibrillation detection. Automated computer-based analysis of continuous ECG telemetry for AF detection in stroke patients had a specificity of 94.6% and sensitivity of 78.4% compared to manual analysis.