Key result
Novel X-ray technology cuts patient radiation dose ~75% during coronary angiography.
RCT (n=70)
Randomly assigned
Effect estimate: 75% reduction
Absolute Event Rate: 12% vs 47%
p-value: p=<0.001
OBJECTIVES: The purpose of this study was to quantify the reduction in patient radiation dose during coronary angiography (CA) by a new X-ray technology, and to assess its impact on diagnostic image quality. BACKGROUND: Recently, a novel X-ray imaging technology has become available for interventional cardiology, using advanced image processing and an optimized acquisition chain for radiation dose reduction. METHODS: 70 adult patients were randomly assigned to a reference X-ray system or the novel X-ray system. Patient demographics were registered and exposure parameters were recorded for each radiation event. Clinical image quality was assessed for both patient groups. RESULTS: With the same angiographic technique and a comparable patient population, the new imaging technology was associated with a 75% reduction in total kerma-area product (KAP) value (decrease from 47 Gycm2 to 12 Gycm2, P<0.001). Clinical image quality showed an equivalent detail and contrast for both imaging systems. On the other hand, the subjective appreciation of noise was more apparent in images of the new image processing system, acquired at lower doses, compared to the reference system. However, the higher noise content did not affect the overall image quality score, which was adequate for diagnosis in both systems. CONCLUSIONS: For the first time, we present a new X-ray imaging technology, combining advanced noise reduction algorithms and an optimized acquisition chain, which reduces patient radiation dose in CA drastically (75%), while maintaining diagnostic image quality. Use of this technology may further improve the radiation safety of cardiac angiography and interventions.
No takes yet. Share an insight, caveat, or question.
Eloot et al. (2015) conducted an RCT in Coronary angiography (n=70). Novel X-ray imaging technology vs. Reference X-ray system was evaluated on Total kerma-area product (KAP) value (75% reduction, p=<0.001). A novel X-ray imaging technology reduced patient radiation dose (total kerma-area product) by 75% (12 vs 47 Gycm2; P<0.001) compared with a reference system during coronary angiography.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: