Key result
Surgical repair is linked to ~45-point lower mortality in Impella CP-associated LV perforation.
Why the study?
Following a 2023 FDA advisory warning of increased LV perforation risk with percutaneous microaxial pumps, this study sought to describe the clinical setting, patterns, high-risk subgroups, and outcomes of LV perforation in real-world reports.
What are the clinical settings, patterns, management, and mortality outcomes of left ventricular perforation associated with Impella microaxial pumps in real-world adverse-event reports?
Observational (n=24,424)
What are the clinical settings, patterns, management, and mortality outcomes of left ventricular perforation associated with Impella microaxial pumps in real-world adverse-event reports?
Left ventricular perforation associated with Impella devices is rare in FDA adverse-event reports (0.77%) but carries a high mortality rate of 30.2%, which is substantially mitigated by intraprocedural recognition and prompt repair.
Answers are grounded in this paper and cite their sources.
Raises perforation awareness during Impella support; leaves open optimal prevention strategies pending prospective data.
BACKGROUND: Percutaneous microaxial pumps are used in cardiogenic shock and high-risk percutaneous coronary intervention (PCI), but left ventricular (LV) perforation is an infrequent but serious complication. In 2023, an FDA advisory warned of increased perforation risk. AIMS: To describe clinical setting, patterns, high-risk subgroups, and outcomes of LV perforation in real-world reports. METHODS: We queried the FDA Manufacturer and User Facility Device Experience (MAUDE) database (2016-2024) for Impella adverse-event reports, screened narratives for perforation terms, and manually adjudicated suspected cases to confirm LV perforation and extract device, clinical setting, presentation, management, and outcomes. RESULTS: Among 24,424 Impella reports, 188 involved LV perforation (0.77%). Impella CP accounted for 52.7% and Impella 5.5 for 41.5% (few with 5.0/2.5; none with RP). Annual LV perforations per 1000 reports declined from 166.7 (2016) to 5.2 (2024). Events occurred most often during cardiac surgery (27.1%), acute coronary syndrome (19.1%), or high-risk PCI (17.0%). Intraprocedural diagnosis occurred in 76.2% and was more common with 5.5 versus CP (83.3% vs. 67.7%). With intraprocedural diagnosis, mortality was 24.6% for 5.5 and 28.4% for CP, versus 66.7% for both when diagnosis was post-procedural. Pericardial effusion (34.9%) and hypotension (28.6%) were common. Repair was attempted at 72.9% and reduced mortality, particularly for CP (19.7% with repair vs. 64.7% without; p < 0.001). Overall mortality was 30.2%, without differences by sex or race. CONCLUSIONS: LV perforation reported with Impella devices is rare but carries high mortality; earlier recognition and prompt repair are associated with improved survival.
No takes yet. Share an insight, caveat, or question.
Rafique et al. (2026) conducted an observational in Impella-associated left ventricular perforation (n=24,424). Impella percutaneous microaxial pump was evaluated on Left ventricular perforation and associated mortality (p=<0.001). Impella-associated LV perforation occurred in 0.77% of MAUDE reports with 30.2% overall mortality, though surgical repair reduced Impella CP mortality.
Synapse has enriched 10 related papers. Consider them for comparative context: