Key result
IEGM-based AV delay optimization correlates well with standard echocardiographic methods.
Why the study?
Atrioventricular delay optimization was used in landmark CRT trials but is often omitted in daily practice, prompting investigation into optimal delays and a simple intracardiac electrogram-based approach.
Does an IEGM-based approach correlate with standard echocardiographic methods for determining optimal atrioventricular delay offsets in patients with a CRT device?
Observational (n=328)
Single-blind (assessors blinded to patient specifics and echocardiographic outcomes)
No
Does an IEGM-based approach correlate with standard echocardiographic methods for determining optimal atrioventricular delay offsets in patients with a CRT device?
An IEGM-based algorithm offers a simple and reliable method to calculate paced AV delay offset, strongly correlating with standard echocardiography, allowing individualized CRT device optimization in clinical practice.
Answers are grounded in this paper and cite their sources.
Supports IEGM-based AV optimization in CRT; leaves open whether it improves outcomes versus echocardiography.
Optimization of the atrioventricular (AV) delay has been performed in several landmark trials in cardiac resynchronization therapy (CRT), although it is often not performed in daily practice. Our aim was to study optimal AV delays and investigate a simple intracardiac electrogram (IEGM)-based optimization approach. 328 CRT patients with paired IEGM and echocardiography optimization data were included in our single-center observational study. Sensed (sAV) and paced (pAV) AV delays were optimized using an iterative echocardiography method. The offset between sAV and pAV delays was calculated using the IEGM method. The mean age of the patients was 69 ± 12 years; 64% were men, 48% had ischemic etiology of heart failure. During echocardiographic optimization, an offset of 73 ± 18 ms was found, differing from nominal AV settings (p < 0.001). Based on the IEGM method, the optimal offset was 75 ± 25 ms. The echocardiographic and IEGM-generated AV offset delays showed good correlation (R2 = 0.62, p < 0.001) and good agreement according to Bland-Altman plot analysis. CRT responders had a near zero offset difference between IEGM and echo optimization (−0.2 ± 17 ms), while non-responders had an offset difference of 6 ± 17 ms, p = 0.006. In conclusion, optimal AV delays are patient-specific and differ from nominal settings. pAV delay can easily be calculated from IEGM after sAV delay optimization.
No takes yet. Share an insight, caveat, or question.
Kloosterman et al. (2023) conducted an observational in Heart failure with cardiac resynchronization therapy (CRT) indication (n=328). Intracardiac electrogram (IEGM)-based AV delay optimization vs. Echocardiographic iterative method and nominal device settings was evaluated on Correlation of optimal offset between sensed and paced AV delays obtained from echocardiography and IEGM methods (R2 = 0.62, p=<0.001). An IEGM-based method for determining optimal paced AV delay offset showed a good correlation with the standard echocardiographic iterative method.
Synapse has enriched 10 related papers. Consider them for comparative context: