Key result
The DRug-Eluting Absorbable Metal Scaffold (DREAMS) demonstrated excellent safety at 3 years with a 6.6% target lesion failure rate and no cardiac death or scaffold thrombosis.
Why the study?
Does a drug-eluting absorbable metal scaffold (DREAMS) provide long-term safety and performance in patients with de novo coronary lesions?
Does a drug-eluting absorbable metal scaffold (DREAMS) provide long-term safety and performance in patients with de novo coronary lesions?
The DREAMS bioresorbable scaffold demonstrated excellent long-term safety at 3 years with low target lesion failure and no scaffold thrombosis in a first-in-man study.
Answers are grounded in this paper and cite their sources.
Supports DREAMS safety in small first-in-man cohort; leaves open randomized efficacy trials before practice change.
Bioresorbable scaffolds were designed to overcome the limitations of permanent stents. In the BIOSOLVE-I study we aimed to assess the long-term safety and performance of a drug-eluting absorbable metal scaffold (DREAMS) at three years.In this prospective, multicentre first-in-man study, 46 patients with 47 de novo lesions were enrolled. We report the final results at three-year follow-up. Mean age was 65.3±9.7 years, lesions were 2.73±0.48 mm in diameter and 10.99±4.59 mm long. Follow-up at three years was available for 44 patients (one patient died of a non-cardiac cause and one patient withdrew consent). Three target lesion failures (TLF) occurred (6.6%), consisting of two clinically driven target lesion revascularisations at scheduled six-month angiography (4.3%) and one myocardial infarction after drug-eluting balloon treatment in a non-target lesion but target vessel at 12-month angiography (2.2%). No cardiac death or scaffold thrombosis occurred. Seven patients had additional angiographic follow-up at 28±4 months: in-scaffold late lumen loss had improved from 0.51±0.46 mm (median 0.28 mm) at 12 months to 0.32±0.32 mm (median 0.20 mm).The BIOSOLVE-I study showed excellent long-term outcomes at three years with a low TLF rate and no cardiac death or scaffold thrombosis. No TLF event was observed beyond 377 days.
No takes yet. Share an insight, caveat, or question.
Haude et al. (2016) studied de novo coronary lesions (n=46). DRug-Eluting Absorbable Metal Scaffold (DREAMS) was evaluated on target lesion failures (TLF). The DRug-Eluting Absorbable Metal Scaffold (DREAMS) demonstrated excellent safety at 3 years with a 6.6% target lesion failure rate and no cardiac death or scaffold thrombosis.