Key result
Left bundle branch area pacing utilizing three-dimensional mapping was successfully performed in all 3 patients, facilitating easy assessment of lead depth.
Why the study?
While left bundle branch area pacing is feasible by advancing the lead deep into the interventricular septum, the utility of three-dimensional mapping to achieve this technique needed to be described.
Does three-dimensional mapping facilitate successful left bundle branch area pacing and lead depth assessment in patients requiring pacing?
Case Report (n=3)
Does three-dimensional mapping facilitate successful left bundle branch area pacing and lead depth assessment in patients requiring pacing?
Three-dimensional mapping is a feasible tool to guide left bundle branch area pacing and assess lead depth, potentially reducing the learning curve for implanters.
Answers are grounded in this paper and cite their sources.
3D mapping may aid LBBP lead placement; hypothesis-generating for procedural optimization, prospective validation needed.
INTRODUCTION: Permanent His bundle pacing is feasible and effective in patients with atrioventricular block or left bundle branch block. However, pacing thresholds to capture the distal His bundle is often higher. Recently left bundle branch area pacing (LBBP) has been shown to be feasible by advancing the lead transvenously, deep into the interventricular septum to reach the left ventricular endocardial surface. In this article we describe the utility of three dimensional (3D) mapping to achieve LBBP. METHODS: Ensite Precision (Abbott) mapping system was used to perform LBBP. A decapolar catheter was used to create 3D map of right atrium and right ventricle (RV). Regions of interest (His bundle, potential LBBP sites of interest in RV) were tagged in the 3D map. The LBBP lead was implanted utilizing the 3D map. The lead depth in the septum was assessed in the 3D map. RESULTS: LBBP was performed in three patients: chronic LBBB and intermittent 2:1 atrioventricular block; atrioventricular (AV) node ablation and conduction system pacing; and bifascicular block and intermittent AV block in a patient with severe left ventricular hypertrophy. LBBP was successful in all three patients. The lead depth in the interventricular septum was 12, 11, and 21 mm, respectively as assessed by 3D mapping. CONCLUSIONS: Three-dimensional mapping was helpful in achieving LBBP in patients with LBBB, severe left ventricular hypertrophy or during AV node ablation. 3D mapping also facilitated easy assessment of lead depth during and after lead fixation. 3D mapping techniques may be a valuable tool to reduce the learning curve of implanters with minimal experience in LBBP.
No takes yet. Share an insight, caveat, or question.
Vijayaraman et al. (2019) conducted a case report in Conduction abnormalities (LBBB, AV block) (n=3). Left bundle branch area pacing (LBBP) utilizing three-dimensional (3D) mapping was evaluated on Procedural success. Left bundle branch area pacing utilizing three-dimensional mapping was successfully performed in all 3 patients, facilitating easy assessment of lead depth.