Key result
A custom pulsatile left ventricular pump and 3D-printed mock circulatory loop successfully reproduced patient-specific aortic and mitral flow and pressure waveforms based on in vivo MRI data.
Why the study?
Mock circulatory loops need enough versatility to reproduce the physiologic range of both flows and pressures for specific fluid dynamic conditions.
A novel, versatile left ventricle pump and 3D-printed mock circulatory loop can accurately reproduce patient-specific aortic hemodynamics in vitro.
Answers are grounded in this paper and cite their sources.
May enhance in vitro cardiac device testing versatility; leaves open in vivo validation before broader adoption.
The importance of experimental setups able to reproduce cardiac functions was well established in the field of clinical innovations. The mock circulatory loops acquired rising relevance, and the possibility to have a complete reproduction of different and specific fluid dynamic conditions within the setup is pivotal. A system with enough versatility to reproduce the physiologic range of both flows and pressures is required. This study describes the design of a versatile setup composed by a custom pulsatile left ventricular pump system and a 3D-printed mock circulatory loop for the in vitro analysis of a patient-specific case of an aortic complex. The performances of the pump were validated first with a set of test flow profiles. It was demonstrated that the system was able to cover a wide range of aortic and mitral flows. Second, the pump system was inserted within the full mock circulatory loop. A patient-specific case was reproduced, both in terms of flow and pressure profiles. A successful validation of the flow and pressure waveforms was obtained by using patient-specific in vivo data from magnetic resonance analysis.
No takes yet. Share an insight, caveat, or question.
Vignali et al. (2022) studied aortic complex. custom pulsatile left ventricular pump system and 3D-printed mock circulatory loop was evaluated on reproduction of flow and pressure profiles. A custom pulsatile left ventricular pump and 3D-printed mock circulatory loop successfully reproduced patient-specific aortic and mitral flow and pressure waveforms based on in vivo MRI data.