Key result
New biventricular circulatory mock loop successfully simulates hemodynamics for healthy and failing hearts.
Why the study?
The study was conducted to evaluate a newly designed circulatory mock loop intended to model cardiac and circulatory hemodynamics for mechanical circulatory support device testing.
A newly developed biventricular circulatory mock loop can reliably simulate hemodynamics of a healthy heart and various heart failure conditions, providing a bench tool for mechanical circulatory support device testing.
May aid preclinical MCS device testing in simulated HF states; leaves open clinical translation.
This study aimed to evaluate a newly designed circulatory mock loop intended to model cardiac and circulatory hemodynamics for mechanical circulatory support device testing. The mock loop was built with dedicated ports suitable for attaching assist devices in various configurations. This biventricular mock loop uses two pneumatic pumps (Abiomed AB5000 ™ , Danvers, MA, USA) driven by a dual-output driver (Thoratec Model 2600, Pleasanton, CA, USA). The drive pressures can be individually modified to simulate a healthy heart and left and/or right heart failure conditions, and variable compliance and fluid volume allow for additional customization. The loop output for a healthy heart was tested at 4.2 L/min with left and right atrial pressures of 1 and 5 mm Hg, respectively; a mean aortic pressure of 93 mm Hg; and pulmonary artery pressure of 17 mm Hg. Under conditions of left heart failure, these values were reduced to 2.1 L/min output, left atrial pressure = 28 mm Hg, right atrial pressure = 3 mm Hg, aortic pressure = 58 mm Hg, and pulmonary artery pressure = 35 mm Hg. Right heart failure resulted in the reverse balance: left atrial pressure = 0 mm Hg, right atrial pressure = 30 mm Hg, aortic pressure = 100 mm Hg, and pulmonary artery pressure = 13 mm Hg with a flow of 3.9 L/min. For biventricular heart failure, flow was decreased to 1.6 L/min, left atrial pressure = 13 mm Hg, right atrial pressure = 13 mm Hg, aortic pressure = 52 mm Hg, and pulmonary artery pressure = 18 mm Hg. This mock loop could become a reliable bench tool to simulate a range of heart failure conditions.
No takes yet. Share an insight, caveat, or question.
Kado et al. (2020) studied Heart failure (simulated). Circulatory mock loop was evaluated on Hemodynamic parameters under various simulated heart conditions. A newly designed biventricular circulatory mock loop successfully simulated hemodynamics for a healthy heart, left heart failure, right heart failure, and biventricular heart failure.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: