Key result
Standard OCT interpretation algorithms are linked to improved fibrous cap thickness reproducibility, raising ICC to ~0.88.
Why the study?
Does the development of standard interpretation algorithms improve the interobserver reproducibility of fibrous cap thickness measurements by OCT in vivo?
Observational (n=170)
Does the development of standard interpretation algorithms improve the interobserver reproducibility of fibrous cap thickness measurements by OCT in vivo?
Answers are grounded in this paper and cite their sources.
Standardized interpretation algorithms significantly improve the interobserver reproducibility of fibrous cap thickness measurements using OCT in vivo.
BACKGROUND: Optical coherence tomography (OCT) imaging is considered to be the only imaging modality with sufficient resolution to measure fibrous cap thickness (FCT) in vivo. However, reproducibility of the measurements in vivo has been unsatisfactory. OBJECTIVES: The authors aimed to investigate whether satisfactory reproducibility of FCT measurements by OCT in vivo can be achieved between independent observers. METHODS: One hundred seventy OCT pullbacks were analyzed by 2 independent observers with intravascular imaging expertise in accordance with current guidelines to assess the interobserver variability of FCT measurement by intraclass correlation coefficient (ICC). The main sources of the variability were analyzed and incorporated in lesion assessment criteria. The same 170 OCT pullbacks were reanalyzed by the same observers using the developed criteria, and the interobserver reproducibility of the measurements was reassessed. On the basis of the developed criteria, a third independent observer interpreted all 170 OCT images. Assessment of the maximal lipid arc was also undertaken similarly before and after the development of interpretation criteria. RESULTS: The original ICC of the FC thickness was 0.56 (95% confidence interval [CI]: 0.38 to 0.69). The poor definition of necrotic core facing border of FC and the neointimal presence of macrophages and calcification contributed to the high interobserver variability of FCT measurement. The ICC of FCT measurements by OCT in vivo was 0.88 (95% CI: 0.80 to 0.93) after we developed lesion assessment criteria. The ICC for the maximal lipid arc assessment before and after was 0.76 and 0.82 respectively. The third independent observer was extensively coached and returned the ICC of 0.82 (95% CI: 0.74 to 0.87) with observer 1 and 0.90 (95% CI: 0.86 to 0.94) with observer 2. CONCLUSIONS: Careful consideration of OCT features mimicking fibroatheroma lesions and imaging artifacts contributed to significantly higher levels of interobserver agreement. Interobserver variation can be partially resolved by development of standard interpretation algorithms.
No takes yet. Share an insight, caveat, or question.
Kini et al. (2016) conducted an observational in Fibroatheroma lesions (n=170). Standard interpretation algorithms for OCT vs. Original interpretation was evaluated on Interobserver variability of FCT measurement by intraclass correlation coefficient (ICC). Development of standard interpretation algorithms for OCT improved the interobserver reproducibility of fibrous cap thickness measurements, increasing the ICC from 0.56 to 0.88.