Key result
CIED extraction achieves similar procedural success in octogenarians compared to younger patients.
Why the study?
Octogenarians present a unique challenge for transvenous CIED extraction due to comorbidities and perceived increased morbidity and mortality, despite proven safety in the general population.
Does transvenous lead extraction of cardiovascular implantable electronic devices have similar efficacy and safety in octogenarians compared to younger patients?
Meta-Analysis (n=4,182)
Does transvenous lead extraction of cardiovascular implantable electronic devices have similar efficacy and safety in octogenarians compared to younger patients?
Transvenous CIED lead extraction can be performed safely and effectively in octogenarians, supporting its use in elderly patients without an elevated risk of major complications or procedural failure.
Answers are grounded in this paper and cite their sources.
Supports CIED extraction in octogenarians; extends evidence of comparable success and safety versus younger patients.
BackgroundTransvenous lead extraction of cardiovascular implantable electronic device (CIED) has been proven safe in the general patient population with the advances in extraction techniques. Octogenarians present a unique challenge given their comorbidities and the perceived increase in morbidity and mortality.ObjectiveTo assess the safety and outcomes of CIED extraction in octogenarians to younger patients.MethodsWe performed an extensive literature search and systematic review of studies that compared CIED extraction in octogenarians versus non-octogenarians. We separately assessed the rate of complete procedure success, clinical success, procedural mortality, major and minor complications. Risk ratio (RR) 95% confidence intervals were measured using the Mantel-Haenszel method. The random effects model was used due to heterogeneity across study cohorts.ResultsSeven studies with a total of 4,182 patients were included. There was no difference between octogenarians and non-octogenarians in complete procedure success (RR 1.01, 95% CI 1.00 – 1.02, p = 0.19) and clinical success (RR 1.01, 95% CI 1.00 – 1.01, p = 0.13). There was also no difference in procedural mortality (RR 1.43, 95% CI 0.46 – 4.39, p = 0.54), major complication (RR 1.40, 95% CI 0.68 – 2.88, p = 0.36), and minor complication (RR 1.43, 95% CI 0.90 – 2.29, p = 0.13).ConclusionIn this study, there was no evidence to suggest a difference in procedural success and complication rates between octogenarians and younger patients. Transvenous lead extraction can be performed safely and effectively in the elderly population. Transvenous lead extraction of cardiovascular implantable electronic device (CIED) has been proven safe in the general patient population with the advances in extraction techniques. Octogenarians present a unique challenge given their comorbidities and the perceived increase in morbidity and mortality. To assess the safety and outcomes of CIED extraction in octogenarians to younger patients. We performed an extensive literature search and systematic review of studies that compared CIED extraction in octogenarians versus non-octogenarians. We separately assessed the rate of complete procedure success, clinical success, procedural mortality, major and minor complications. Risk ratio (RR) 95% confidence intervals were measured using the Mantel-Haenszel method. The random effects model was used due to heterogeneity across study cohorts. Seven studies with a total of 4,182 patients were included. There was no difference between octogenarians and non-octogenarians in complete procedure success (RR 1.01, 95% CI 1.00 – 1.02, p = 0.19) and clinical success (RR 1.01, 95% CI 1.00 – 1.01, p = 0.13). There was also no difference in procedural mortality (RR 1.43, 95% CI 0.46 – 4.39, p = 0.54), major complication (RR 1.40, 95% CI 0.68 – 2.88, p = 0.36), and minor complication (RR 1.43, 95% CI 0.90 – 2.29, p = 0.13). In this study, there was no evidence to suggest a difference in procedural success and complication rates between octogenarians and younger patients. Transvenous lead extraction can be performed safely and effectively in the elderly population.
No takes yet. Share an insight, caveat, or question.
Lin et al. (2020) conducted a meta-analysis in Cardiovascular implantable electronic device (CIED) extraction (n=4,182). Age ≥80 years (octogenarians) undergoing CIED extraction vs. Age <80 years (younger patients) undergoing CIED extraction was evaluated on Complete procedure success (RR 1.01, 95% CI 1.00-1.02, p=0.19). Cardiovascular implantable electronic device extraction had similar rates of complete procedure success in octogenarians compared to younger patients (RR 1.01).