Key result
Elevated D-dimer levels during venovenous ECMO reflected clot formation within the membrane oxygenator (clot volume 16-106 cm3), with D-dimers decreasing after oxygenator exchange.
Why the study?
Do elevated D-dimer levels predict clot volume inside membrane oxygenators during venovenous ECMO?
Observational (n=13)
Do elevated D-dimer levels predict clot volume inside membrane oxygenators during venovenous ECMO?
Elevated D-dimer levels during venovenous ECMO reflect coagulation activity and predict clot formation within the membrane oxygenator, highlighting the importance of routine D-dimer monitoring to prevent system-induced coagulation disorders.
Answers are grounded in this paper and cite their sources.
Elevated D-dimers signal MO thrombosis during vv-ECMO; leaves open optimal monitoring thresholds and exchange timing.
Thrombosis inside the membrane oxygenator (MO) is a critical complication during venovenous extracorporeal membrane oxygenation (ECMO). The aim of this study was to prove if thrombotic clots manifest within the MO when D-dimer levels are elevated over a long-term period. Heparin-coated polymethylpentene MOs (n = 13) were exchanged due to high plasma D-dimer levels. Clot volume was calculated using multidetector computed tomography (MDCT). Coagulation parameters and MO function were analyzed before and after MO exchange. Before MO exchange, D-dimer levels increased significantly in each patient (11.5 [6.5-15.5] mg/L to 35.0 [34-35] mg/L, P ≤ 0.001). High levels of D-dimers were tolerated for 1 to 6 days. Additionally, fibrinogen concentration (n = 8) and platelet count decreased (n = 8). Within 48 h after exchange, D-dimer levels decreased significantly (n = 11, 12 [8-16] mg/L, P = 0.004). Fibrinogen concentration and platelet counts increased. Clots were found in all MOs in the inlet part of the device. Clot volume (16-106 cm(3) ) did not correlate with MO support time but increased significantly when high D-dimer levels were accepted for >2 days. An increase or high levels of D-dimers in absence of other explaining pathology during ECMO therapy reflected coagulation activity within the MO. Evidence of clots within the MO at high D-dimer levels and decrease after exchange underline the relevance of D-dimer testing during ECMO treatment. Besides, surveillance of MOs during ongoing ECMO therapy will help to predict clot formation, and to avoid system-induced coagulation disorders as well as critical situations.
No takes yet. Share an insight, caveat, or question.
Dornia et al. (2015) conducted an observational in Venovenous extracorporeal membrane oxygenation (ECMO) (n=13). Elevated D-dimer levels vs. Baseline or post-exchange D-dimer levels was evaluated on Clot volume inside membrane oxygenators and D-dimer levels before and after exchange (p=≤0.001). Elevated D-dimer levels during venovenous ECMO reflected clot formation within the membrane oxygenator (clot volume 16-106 cm3), with D-dimers decreasing after oxygenator exchange.