Key result
Polymethyl-pentene hollow fiber oxygenators yield 100% survival without device failure in adult ECLS cases.
Case Report (n=6)
Answers are grounded in this paper and cite their sources.
Silicone oxygenators are the standard devices used for Extracorporeal Life Support (ECLS), but they have some limitations. Microporous polypropylene hollow fiber oxygenators overcome many of these problems but, unfortunately, develop plasma leak. Polymethyl-pentene (PMP) is a novel oxygenator material. We report our initial experience with the Medos Hilite 7000LT, a PMP hollow fiber oxygenator, in six adult respiratory ECLS patients with these characteristics: age, mean 32.2 (+/-13) years; weight, mean 81.2 (+/-17) kg; PaO2/FIO2, mean 62.8 [+/-33] mm Hg; Murray Score, mean 3.4 [+/-0.3]; and sepsis related organ failure assessment score, mean 9.6 [+/-2.3]. One patient was cannulated within 10 hours of multiple trauma and 1 hour after thoracolaparotomy; another patient was cannulated 12 hours after a thoracotomy. All six patients survived. Heparin was infused (7.8-32.5 u/kg/hr) to maintain activated clotting time at 162 to 238 seconds; international normalized ratio was 0.9 to 3.4. Two of the six patients required transfusions of fresh frozen plasma, receiving one and five units, respectively. Fibrinogen was 1.4 to 6 g/dl; no cryoprecipitate was needed. Platelet counts were between 65 and 306, and very little platelet transfusion (mean 2.33; +/-3.03 units per patient) was required to maintain these levels. Two patients did not require any platelet transfusion. Maximum blood flow was 5.3 L/min, sweep was 3 to 10 L/min, and resistance was 11 to 43 Paul Wood Units. There were no oxygenator failures. Mean duration of ECLS was 151.7 hours (+/-75.6). Our initial experience with PMP oxygenators in adults was satisfactory, and platelet consumption and resistance to blood flow seem to be greatly reduced with PMP.
No takes yet. Share an insight, caveat, or question.
Peek et al. (2002) conducted a case report in Respiratory failure requiring ECLS (n=6). Polymethyl-pentene (PMP) hollow fiber oxygenator (Medos Hilite 7000LT) was evaluated on Survival. The use of a polymethyl-pentene hollow fiber oxygenator in six adult ECLS patients resulted in 100% survival, no oxygenator failures, and apparently reduced platelet consumption.
Synapse has enriched 10 related papers. Consider them for comparative context: