Answers are grounded in this paper and cite their sources.
Neural implants allow to decipher brain functions as well as to design brain–computer interfaces that aim to compensate for the loss of functions after a brain injury. After two decades of development worldwide, the efficiency of the recording/stimulating implants is now sufficient to consider their transfer to clinical applications. Nevertheless, this translation is slowed down by a lack of proof of long‐term efficacy. Consequently, strong research effort is currently dedicated to obtain devices that can record and stimulate for decades. The major sources of failure that are identified are the delamination of the insulation and contact layers of the implant as well as the formation of cracks in the substrate. One of the main challenges is thus to find novel highly stable substrates and insulators. Diamond is an interesting candidate to meet this requirement, offering both high biocompatibility and resistance to corrosion. Herein, thin diamond layers are combined with polyimide substrate to form the backbone of a neural implant and its functionality is assessed. The microelectrodes are made of titanium nitride on platinum coated with poly(3,4‐ethylenedioxythiophene):poly(styrene sulfonate). This new type of neural implant is successfully tested in vivo to measure cortical auditory‐evoked activity in rats.
No takes yet. Share an insight, caveat, or question.
Hébert et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: