Key result
Blood anticoagulated with hirudin yielded complete blood counts and clinical chemistry results that were highly predictable from routine measurements.
Why the study?
Does hirudin serve as an effective universal anticoagulant for automated complete blood counts, clinical chemistry, and immunohaematology compared to standard processing?
Cross-Sectional (n=80)
Does hirudin serve as an effective universal anticoagulant for automated complete blood counts, clinical chemistry, and immunohaematology compared to standard processing?
Hirudin at 400 ATU/ml effectively anticoagulates blood for diagnostic purposes, potentially allowing a single universal blood sampling tube for haematology, clinical chemistry, and blood grouping.
Answers are grounded in this paper and cite their sources.
May support hirudin as alternative anticoagulant; leaves open routine adoption pending larger validation.
Undesirable interactions between anticoagulants and diagnostic test kit procedures so far have prevented the development of a single uniform blood sampling tube. Contrary to K2-EDTA, heparin and other anticoagulants, hirudin only minimally alters blood cells and dissolved blood constituents, thus qualifying as a universal anticoagulant for diagnostic purposes. Automated complete blood counts, automated analyses of clinical chemistry analytes and immunohaematology were performed from hirudinised and routinely processed blood obtained from healthy volunteers (n=35) and hospitalised patients (n=45). Hirudin (400 ATU/ml blood) sufficiently anticoagulated blood for diagnostic purposes. The measurements of automated complete blood counts obtained from K2-EDTA-anticoagulated and hirudinised blood correlated significantly as did the measurements of 24 clinical chemistry analytes from hirudinised plasma and serum. Regression analysis revealed that the results of complete blood counts and clinical chemistry tests were predictable from the respective measurements from hirudinised blood (p=0.001). Immunohaematological tests and cross-matching from hirudinised and native blood of the same donors gave identical results. Single clotting factors, but not global coagulation analytes, could be measured from hirudinised blood. Therefore, a universal hirudin-containing blood sampling tube could be designed for automated analysis of haematological, serological and clinical chemistry analytes.
No takes yet. Share an insight, caveat, or question.
Menssen et al. (2001) conducted a cross-sectional in Diagnostic blood sampling (n=80). Hirudin vs. K2-EDTA and routinely processed blood (serum, native blood) was evaluated on Correlation of automated complete blood counts and clinical chemistry analytes between hirudinised and routinely processed blood (p=0.001). Blood anticoagulated with hirudin yielded complete blood counts and clinical chemistry results that were highly predictable from routine measurements.