Key result
3D-TEE guides successful femoral snare extraction of a retained ICD lead after conventional failure.
Why the study?
Transvenous pacer and defibrillator lead extraction is standard for CIED infection but carries significant procedural risk, particularly with advanced cardiomyopathy, active infection, and long lead dwell times.
Case Report (n=1)
No
3D-TEE can provide critical real-time spatial guidance to facilitate successful transvenous lead extraction when conventional fluoroscopic techniques fail in high-risk patients.
May aid complex lead extraction when fluoroscopy fails; hypothesis-generating case report requiring prospective validation.
Transvenous pacer and defibrillator lead extraction is the standard treatment for cardiac implantable electronic device (CIED) infection but carries significant procedural risk, particularly in patients with advanced cardiomyopathy, active infection, and long-standing lead dwell times. A 54-year-old male with nonischemic cardiomyopathy (ejection fraction <20%) and end-stage renal disease on hemodialysis presented with methicillin-resistant Staphylococcus aureus (MRSA) bacteremia in the setting of a biventricular implantable cardioverter-defibrillator (ICD) placed 13 years earlier. Echocardiography demonstrated severe systolic dysfunction and a mobile echodensity consistent with vegetation on a pacemaker lead near the superior vena cava-right atrial junction, necessitating complete device extraction. Initial laser-assisted extraction in the operating room under general anesthesia with invasive monitoring and transesophageal echocardiography (TEE) guidance successfully removed one lead, but a second lead could not be fully extracted despite attempted removal. The patient was maintained on targeted intravenous antibiotics, and completion extraction was performed two weeks later; however, multiple conventional snaring attempts were unsuccessful due to lead orientation and adherence within the right atrium. Three-dimensional TEE (3D-TEE) was subsequently used to improve spatial visualization, enabling successful femoral snare capture and removal of the retained lead without complication. This case highlights the importance of procedural planning and advanced imaging in high-risk lead extraction, as a staged approach may mitigate risk while 3D-TEE provides critical real-time spatial guidance when conventional techniques fail.
No takes yet. Share an insight, caveat, or question.
Sikka et al. (2026) conducted a case report in Cardiac implantable electronic device (CIED) infection (n=1). 3D transesophageal echocardiography (3D-TEE) guided transvenous lead extraction was evaluated on Successful lead extraction. Three-dimensional transesophageal echocardiography successfully guided the femoral snare capture and removal of a retained ICD lead in a high-risk patient after conventional techniques failed.