Key result
Evolution Mechanical Dilator Sheath achieves complete extraction of an ICD lead with large vegetation without complications.
Why the study?
Surgical removal of large vegetations on ICD leads can be highly risky, and alternative percutaneous extraction methods need evaluation.
Case Report (n=1)
May support mechanical extraction in select high-risk cases with vegetations; hypothesis-generating and should not yet change practice.
A 74-year-old male patient presented with complaints of fatigue, fever, and worsening health status. He had a 34-month history of cardioverter-defibrillator (ICD) implantation due to ischemic cardiomyopathy. Transthoracic and transesophageal echocardiography showed a mobile mass, 24 x 11 mm in size, consistent with a vegetation attached to the right ventricular ICD lead. Surgical removal of the electrode was considered to be highly risky for the patient, thus percutaneous removal was decided. Due to adhesions, manual traction of the lead with a standard stylet was ineffective. Complete lead extraction was accomplished using the Evolution Mechanical Dilator Sheath, which is a new mechanical sheath with a stainless steel bladed tip. No complications occurred during or after the procedure. Following a histopathologic diagnosis of vegetation, the patient received a six-week antibiotic therapy, after which a new ICD was implanted on the contralateral side. To our knowledge, this case represents the largest lead vegetation extracted by this new device. Although its safety should be validated by increasing number of cases, it seems that its simple use would make this device a new interesting tool among the instruments available for lead extraction.
No takes yet. Share an insight, caveat, or question.
Tolga Aksu (2011) conducted a case report in ICD lead vegetation (n=1). Evolution Mechanical Dilator Sheath was evaluated on Complete lead extraction and procedural complications. The Evolution Mechanical Dilator Sheath successfully accomplished complete extraction of an ICD lead with a 24 x 11 mm vegetation without complications in a high-surgical-risk patient.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: