Key result
Home self-measurement with a wrist-cuff device was comparable to 24-hour ambulatory monitoring in detecting white-coat hypertension and nocturnal non-dipping in diabetic patients.
Why the study?
Does home self measurement of blood pressure compare to 24-hour ABPM for detecting white coat hypertension and dipping patterns in diabetic patients with uncontrolled hypertension?
RCT (n=78)
Open-label
Randomly assigned
No
Does home self measurement of blood pressure compare to 24-hour ABPM for detecting white coat hypertension and dipping patterns in diabetic patients with uncontrolled hypertension?
Answers are grounded in this paper and cite their sources.
Home self-measurement of blood pressure is comparable to 24-hour ABPM for detecting white coat hypertension and dipping patterns in diabetic patients, offering a reliable alternative for risk assessment.
Background: In modern epidemiology, risk assessment is a crucial step in diabetes care. Clinic blood pressure reading though is not a good measurement for this purpose since both uncontrolled hypertension and white coat hypertension (WCH) are frequent among patients with diabetes mellitus (DM). Given the problems with clinical application of ambulatory blood pressure monitoring (ABPM), in this study we evaluated clinical utility of home self measurement (HSM) with a wrist-cuff device in DM patients with hypertension to make a BP profile. Also, the clinical application of a complex of arterial study, albuminuria and blood pressure profile, in DM risk assessment was investigated. Methods and Materials: Seventy-eight adult DM patients with labile or uncontrolled hypertension were randomly assigned to 24 hour ABPM or HSM for 4 consecutive days and their BP profiles were evaluated in conjunction with an assessment of arterial stiffness and renal function as well as lipid profile. Results: The two groups were of comparable age, gender, BP, DM duration and control, smoking, lipids, renal function, arterial compliance and antihypertensive medication use. ABPM detected 33% WCH and 17.6% evening/night-time dipping, compared to 32% and 16% respectively for HSM, with overlapping 95% confidence intervals for day versus night BP regression coefficients. WCH patients had more compliant arteries as well as less albuminuria compared to the sustained hypertensive group. Conclusion: A complex of BP profile (by either ABPM or HSM), arterial compliance and albuminuria is a reliable and economical alternative to current methods for risk assessment in hypertensive diabetic patients.
No takes yet. Share an insight, caveat, or question.
Alireza Khoshdel (2011) conducted an RCT in Diabetes Mellitus with labile or uncontrolled hypertension (n=78). Home self-measurement (HSM) with a wrist-cuff device vs. 24-hour ambulatory blood pressure monitoring (ABPM) was evaluated on Detection of white-coat hypertension (WCH) (p=1.00). Home self-measurement with a wrist-cuff device was comparable to 24-hour ambulatory monitoring in detecting white-coat hypertension and nocturnal non-dipping in diabetic patients.
Synapse has enriched 10 related papers. Consider them for comparative context: