Key result
Two pediatric patients with congenital complete heart block experienced late dislodgement of Medtronic SelectSecure 3830 pacing leads at 8.8 and 11.8 years post-implantation due to patient growth.
Why the study?
A potential long-term pitfall of sturdy SelectSecure lumenless 3830 pacing leads is relative retraction with somatic growth in children resulting in late lead dislodgement.
Case Report (n=2)
Growing pediatric patients with transvenous pacing leads require frequent radiographic surveillance for lead tautness to prevent late dislodgement.
Answers are grounded in this paper and cite their sources.
Late dislodgement of SelectSecure 3830 leads may occur after 11 years in children; case reports leave open optimal lead selection and surveillance duration.
BACKGROUND: The SelectSecure lumenless 3830 pacing lead is often considered to be the pacing lead of choice for transvenous pacing in children because of its small diameter, lead strength, and reliable long-term sensing and pacing characteristics. One of the potential long-term pitfalls of a sturdy pacing lead is relative retraction with growth in children resulting in late lead dislodgement. CASE SUMMARY: We report two cases of late SelectSecure 3830 lead dislodgement at 11.8 years (Case 1) and 8.8 years (Case 2), respectively, post the initial implantation. Case 1 was diagnosed with congenital complete heart block (CHB) at 9 months old when he presented with unconfirmed diphtheria infection. Case 2 was diagnosed with CHB at 14 weeks of age with positive maternal anti-Ro antibodies. Both patients underwent implantation of a transvenous permanent pacemaker implantation with Medtronic SelectSecure 3830 lead due to symptomatic bradycardia. Apart from a pulse generator change at 8.5 years (Case 1) and 7 years (Case 2), respectively, post-implant due to normal battery depletion, both patients are well in the interim. DISCUSSION: As part of the pacemaker follow-up for rapidly growing children, we recommend more frequent surveillance of lead 'tautness' by chest radiography especially in children with CHB with no underlying heart rhythm.
No takes yet. Share an insight, caveat, or question.
Ng et al. (2021) conducted a case report in Congenital complete heart block (CHB) (n=2). Medtronic SelectSecure 3830 pacing lead was evaluated on Late lead dislodgement. Two pediatric patients with congenital complete heart block experienced late dislodgement of Medtronic SelectSecure 3830 pacing leads at 8.8 and 11.8 years post-implantation due to patient growth.
Synapse has enriched 10 related papers. Consider them for comparative context: