Key result
Transvenous pacemaker implantation was feasible and safe in 12 children weighing <10 kg, with no procedural complications, one lead dislodgment, and one death over a mean 31.8-month follow-up.
Why the study?
Is transvenous pacemaker implantation feasible and safe in children weighing less than 10 kg?
Observational (n=12)
Is transvenous pacemaker implantation feasible and safe in children weighing less than 10 kg?
Transvenous pacemaker implantation is a feasible and safe, less invasive alternative to the standard epicardial approach in children weighing under 10 kg.
Answers are grounded in this paper and cite their sources.
May warrant further study in select infants; leaves open comparative safety versus epicardial pacing.
Pacemakers are used in small children with increasing frequency for the treatment of life-threatening bradyarrhythmias. The epicardial approach is generally preferred in these patients, to avoid the risks of vessel thrombosis. We examined the feasibility and safety of transvenous pacemaker implantation in children weighing <10 kg, via subclavian puncture, using a 4 Fr sheath introduced after a venogram was performed to evaluate the vein diameter. Progressive dilation with 5, 6, and 7 Fr sheaths preceded the insertion and placement of the endocardial lead. A subaponeurotic pocket was created in the abdominal or pectoral regions, depending upon the patient's size. Between 2001 and 2007, we treated 12 patients (median age = 16 months; range 1-32; median weight = 7.9 kg; range 2.3-10.0; 7 males), of whom four weighed <5 kg. Indications for permanent pacing included postsurgical complete atrioventricular block (n = 8), sinus node dysfunction (n = 2), congenital atrioventricular block (n = 1), and long QT syndrome (n = 1). Single-chamber pacemakers were implanted in 10, and dual-chamber pacemakers in two patients. The patients were evaluated at 48 hours, 10 days, and at 3 and 6 months. The mean follow-up was 31.8 +/- 23.5 months. There were no procedural complications. Lead dislodgment occurred in one patient and required replacement of the ventricular lead. One patient died from septicemia. Endocardial pacemaker implantation was feasible and safe in children weighing <10 kg. This procedure is less invasive than the standard epicardial approach.
No takes yet. Share an insight, caveat, or question.
Robledo‐Nolasco et al. (2009) conducted an observational in Life-threatening bradyarrhythmias (n=12). Transvenous pacemaker implantation was evaluated on Procedural complications. Transvenous pacemaker implantation was feasible and safe in 12 children weighing <10 kg, with no procedural complications, one lead dislodgment, and one death over a mean 31.8-month follow-up.
Synapse has enriched 10 related papers. Consider them for comparative context: