Key result
Undergoing 1.5 Tesla MRI in patients with Evia pacemakers and Safio S leads was feasible and safe, resulting in a 100% serious adverse device effect-free rate and no clinically relevant changes in device function.
Why the study?
Is MRI at 1.5 Tesla safe and feasible in patients with the Evia pacemaker and Safio S leads?
Observational (n=30)
None
None
No
Is MRI at 1.5 Tesla safe and feasible in patients with the Evia pacemaker and Safio S leads?
The use of the MR conditional Evia pacemaker system with Safio S leads is safe and feasible under 1.5 Tesla MRI conditions without clinically relevant changes in device function.
Answers are grounded in this paper and cite their sources.
Supports MR-conditional PM feasibility at 1.5T in pilot; leaves open larger confirmatory trials before routine use.
BACKGROUND: The purpose of this study was to evaluate the feasibility of the magnetic resonance (MR) conditional pacemaker (PM) system (Evia SR-T and DR-T with Safio S leads) under MR conditions. METHODS: Patients with standard PM indications and Evia PM were eligible for enrollment in this single center prospective non-randomized pilot study. Patients underwent MR of the brain and lower lumbar spine at 1.5 Tesla. Atrial (RA) und ventricular (RV) lead parameters (sensing, pacing threshold [PTH], pacing impedance) were assessed immediately before (baseline follow-up [FU]) and immediately after MRI (1st FU), after 1 month (2nd FU) and 3 months (3rd FU). The effect of MR on serious adverse device effect (SADE) free-rate, on atrial and ventricular sensing (AS/VS; mV) and atrial (RA) and ventricular (RV) pacing thresholds (PTH; V/0.4 ms) were investigated between baseline and 2nd FU. Continuous variables are expressed as mean ± SD and were compared using paired Student's t-test. A p < 0.05 was considered significant. RESULTS: Thirty-one patients were enrolled. One patient had to be excluded because of an enrollment violation. Therefore, data of 30 patients (female 12 [40%], age 73 ± 12 years, dual chamber PM 15 [50%]) were included in this analysis. No MR related SADE occurred. Lead measurements were not statistically different between the baseline FU and the 2nd FU (AS/VS at baseline 3.2 ± 2.1/15.0 ± 6.0, at 2nd FU 3.2 ± 2.1/14.9 ± 6.5; p = ns. RA-PTH/RV-PTH at baseline 0.68 ± 0.18/0.78 ± 0.22, at 2nd FU 0.71 ± 0.24/0.78 ± 0.22; p = ns). The presence of the permanent pacemakers led to MR imaging artifacts on diffusion weighted sequences of the brain, but did not affect other sequences (e.g. FLAIR and T2 weighted spin-echo images). CONCLUSION: The use of the MR conditional Evia PM in a MR environment under predefined conditions is feasible. No MR related SADEs nor clinically relevant changes in device functions occurred.
No takes yet. Share an insight, caveat, or question.
Wollmann et al. (2012) conducted an observational in Patients with standard pacemaker indications (n=30). 1.5 Tesla MRI scan was evaluated on MRI and pacing system related serious adverse device effect (SADE) free-rate. Undergoing 1.5 Tesla MRI in patients with Evia pacemakers and Safio S leads was feasible and safe, resulting in a 100% serious adverse device effect-free rate and no clinically relevant changes in device function.
Synapse has enriched 10 related papers. Consider them for comparative context: